Option
Description
Hold
The sensor output value is held at the current measured value
during the calibration procedure.
Transfer
A preset output value is sent during calibration. Refer to the
controller user manual to change the preset value.
8.
With the sensor in the reference solution, push
ENTER
.
9.
Enter the reference temperature of the reference solution and push
ENTER
.
10.
Enter the slope of the reference solution and push
ENTER
.
11.
Wait for the value to stabilize and push
ENTER
.
Note: The screen may advance to the next step automatically.
12.
Use the arrow keys to enter the value of the reference solution and
push
ENTER
.
13.
Review the calibration result:
• Passed—the sensor is calibrated and ready to measure samples.
The slope and/or offset values are shown.
• Failed—the calibration slope or offset is outside of accepted limits.
Repeat the calibration with fresh reference solutions. Refer to
on page 15 and
on page 16 for
more information.
14.
If the calibration passed, push
ENTER
to continue.
15.
If the option for operator ID is set to Yes in the Calibration Options
menu, enter an operator ID. Refer to
on page 14.
16.
On the New Sensor screen, select whether the sensor is new:
Option Description
Yes
The sensor was not calibrated previously with this controller. The
days of operation and previous calibration curves for the sensor are
reset.
No
The sensor was calibrated previously with this controller.
17.
Return the sensor to the process and push
ENTER
.
The output signal returns to the active state and the measured
sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when
the outputs return to the active state.
Calibration with the process sample
The sensor can remain in the process sample, or a portion of the
process sample can be removed for calibration.
1.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate.
2.
Select the calibration for the specified parameter and push
ENTER
:
• Conductivity—Cond Cal
• TDS—TDS Cal
• Salinity—Cond Cal
• Concentration—Conc Cal or Cond Cal
3.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
4.
Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during
the calibration procedure.
Hold
The sensor output value is held at the current measured value
during the calibration procedure.
Transfer
A preset output value is sent during calibration. Refer to the
controller user manual to change the preset value.
5.
With the sensor in the process sample, push
ENTER
.
The measured value is shown.
6.
Wait for the value to stabilize and push
ENTER
.
7.
Use the arrow keys to enter the value of the process sample and
push
ENTER
.
English
13
Summary of Contents for Polymetron Series
Page 8: ...1 2 3 4 8 English...
Page 9: ...5 6 7 8 English 9...
Page 26: ...1 2 3 4 26 Deutsch...
Page 27: ...5 6 7 8 Deutsch 27...
Page 46: ...1 2 3 4 46 Italiano...
Page 47: ...5 6 7 8 Italiano 47...
Page 64: ...1 2 3 4 64 Fran ais...
Page 65: ...5 6 7 8 Fran ais 65...
Page 81: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT Espa ol 81...
Page 83: ...1 2 3 4 Espa ol 83...
Page 84: ...5 6 7 8 84 Espa ol...
Page 103: ...1 2 3 4 Portugu s 103...
Page 104: ...5 6 7 8 104 Portugu s...
Page 121: ...1 2 3 4 e tina 121...
Page 122: ...5 6 7 8 122 e tina...
Page 139: ...1 2 3 4 Dansk 139...
Page 140: ...5 6 7 8 140 Dansk...
Page 158: ...1 2 3 4 158 Nederlands...
Page 159: ...5 6 7 8 Nederlands 159...
Page 178: ...1 2 3 4 178 Polski...
Page 179: ...5 6 7 8 Polski 179...
Page 196: ...1 2 3 4 196 Svenska...
Page 197: ...5 6 7 8 Svenska 197...
Page 214: ...1 2 3 4 214 Suomi...
Page 215: ...5 6 7 8 Suomi 215...
Page 230: ...2 3 236 2 1 3 2 4 230...
Page 231: ...3 1 Flow thru T PVC 3 4 NPT 2 Flow thru T 1 4 NPT 231...
Page 232: ...3 3 Polymetron No 1 2 3 4 5 6 7 8 9 10 11 12 232...
Page 233: ...1 2 3 4 233...
Page 234: ...5 6 7 8 234...
Page 247: ...2 ISO 7888 ASTM D5391 64 2 52 08394 A 2000 FDA EN 10204 3 1 B 0 4 m 64 2 52 08394 A 2011 247...
Page 254: ...1 2 3 4 254 Magyar...
Page 255: ...5 6 7 8 Magyar 255...
Page 273: ...1 2 3 4 Rom n 273...
Page 274: ...5 6 7 8 274 Rom n...
Page 292: ...1 2 3 4 292 lietuvi kalba...
Page 293: ...5 6 7 8 lietuvi kalba 293...
Page 308: ...2 3 314 2 1 3 2 4 308...
Page 309: ...3 1 3 4 NPT 2 1 4 NPT 309...
Page 310: ...3 3 Polymetron 1 2 3 4 5 6 7 8 9 10 11 12 310...
Page 311: ...1 2 3 4 311...
Page 312: ...5 6 7 8 312...
Page 315: ...1 2 3 4 5 6 7 8 9 sc100 12 10 318 11 12 1 2 4 315...
Page 316: ...4 3 30 4 5 6 7 8 9 10 11 12 13 319 319 14 15 318 316...
Page 317: ...16 17 1 2 3 4 5 6 7 8 319 319 9 10 318 11 12 317...
Page 318: ...1 2 3 4 1 5 6 7 1 2 1 2 1 7 30 60 90 6 9 1 2 1 2 318...
Page 319: ...3 4 5 318 6 7 Modbus Modbus MSDS SDS 1 2 2 3 3 4 5 5 5 2 3 6 319...
Page 320: ...319 1 2 4 4 1089 1106 23 27 C 5 5 5 5 Sensor Days 320...
Page 322: ...7 100 C TEMP TOO LOW T 0 C 365 320 7 0 4 C 0 200 C 8 8 322...
Page 330: ...1 2 3 4 330 T rk e...
Page 331: ...5 6 7 8 T rk e 331...
Page 349: ...1 2 3 4 Slovensk jazyk 349...
Page 350: ...5 6 7 8 350 Slovensk jazyk...
Page 368: ...1 2 3 4 368 Slovenski...
Page 369: ...5 6 7 8 Slovenski 369...
Page 387: ...1 2 3 4 Hrvatski 387...
Page 388: ...5 6 7 8 388 Hrvatski...
Page 403: ...2 3 409 2 1 3 2 4 403...
Page 404: ...3 1 PVC 3 4 in NPT 2 1 4 in NPT 404...
Page 405: ...3 3 Polymetron 1 2 3 4 5 6 7 8 9 10 11 12 405...
Page 406: ...1 2 3 4 406...
Page 407: ...5 6 7 8 407...
Page 415: ...Borax 1 2 2 3 3 4 5 HCl 5 5 2 3 6 414 1 2 4 4 1089 1106 ohm 23 27 C 5 ohm 5 ohm 415...
Page 418: ...7 HOLD TC 0 4 C 0 200 C TC MENU 8 8 Event OK 8 Event K ADC ADC FLASH ERASE Flash 20 200 C 418...
Page 425: ...1 2 3 4 eesti keel 425...
Page 426: ...5 6 7 8 426 eesti keel...
Page 439: ...ESD 1 439...
Page 441: ...3 1 T PVC 3 4 NPT 2 T 1 4 NPT 441...
Page 442: ...3 3 Polymetron 1 2 3 Polymetron 3 4 5 6 7 8 9 10 11 12 442...
Page 443: ...1 2 3 4 443...
Page 444: ...5 6 7 8 444...
Page 454: ......
Page 455: ......