11.
On the New Sensor screen, select whether the sensor is new:
Option Description
Yes
The sensor was not calibrated previously with this controller. The
days of operation and previous calibration curves for the sensor are
reset.
No
The sensor was calibrated previously with this controller.
12.
Proceed to the calibration with a reference solution or process
sample.
Calibration with a reference solution
Calibration adjusts the sensor reading to match the value of a reference
solution. Use a reference solution that is at the same value or higher
than the expected measurement readings.
Note: If the sensor is being calibrated for the first time, be sure to complete the
zero calibration first.
1.
Thoroughly rinse the clean sensor in deionized water.
2.
Put the sensor in the reference solution. Support the sensor so that it
does not touch the container. Make sure that the sensing area is fully
immersed in the solution (
). Stir the sensor to remove
bubbles.
Figure 4 Sensor in reference solution
3.
Wait for the sensor and solution temperature to equalize. This can
take 30 minutes or more if the temperature difference between the
process and reference solution is significant.
4.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate.
5.
Select the calibration for the specified parameter and push
ENTER
:
• Conductivity—Cond Cal
• TDS—TDS Cal
• Salinity—Cond Cal
• Concentration—Conc Cal or Cond Cal
6.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
7.
Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during
the calibration procedure.
12
English
Summary of Contents for Polymetron Series
Page 8: ...1 2 3 4 8 English...
Page 9: ...5 6 7 8 English 9...
Page 26: ...1 2 3 4 26 Deutsch...
Page 27: ...5 6 7 8 Deutsch 27...
Page 46: ...1 2 3 4 46 Italiano...
Page 47: ...5 6 7 8 Italiano 47...
Page 64: ...1 2 3 4 64 Fran ais...
Page 65: ...5 6 7 8 Fran ais 65...
Page 81: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT Espa ol 81...
Page 83: ...1 2 3 4 Espa ol 83...
Page 84: ...5 6 7 8 84 Espa ol...
Page 103: ...1 2 3 4 Portugu s 103...
Page 104: ...5 6 7 8 104 Portugu s...
Page 121: ...1 2 3 4 e tina 121...
Page 122: ...5 6 7 8 122 e tina...
Page 139: ...1 2 3 4 Dansk 139...
Page 140: ...5 6 7 8 140 Dansk...
Page 158: ...1 2 3 4 158 Nederlands...
Page 159: ...5 6 7 8 Nederlands 159...
Page 178: ...1 2 3 4 178 Polski...
Page 179: ...5 6 7 8 Polski 179...
Page 196: ...1 2 3 4 196 Svenska...
Page 197: ...5 6 7 8 Svenska 197...
Page 214: ...1 2 3 4 214 Suomi...
Page 215: ...5 6 7 8 Suomi 215...
Page 230: ...2 3 236 2 1 3 2 4 230...
Page 231: ...3 1 Flow thru T PVC 3 4 NPT 2 Flow thru T 1 4 NPT 231...
Page 232: ...3 3 Polymetron No 1 2 3 4 5 6 7 8 9 10 11 12 232...
Page 233: ...1 2 3 4 233...
Page 234: ...5 6 7 8 234...
Page 247: ...2 ISO 7888 ASTM D5391 64 2 52 08394 A 2000 FDA EN 10204 3 1 B 0 4 m 64 2 52 08394 A 2011 247...
Page 254: ...1 2 3 4 254 Magyar...
Page 255: ...5 6 7 8 Magyar 255...
Page 273: ...1 2 3 4 Rom n 273...
Page 274: ...5 6 7 8 274 Rom n...
Page 292: ...1 2 3 4 292 lietuvi kalba...
Page 293: ...5 6 7 8 lietuvi kalba 293...
Page 308: ...2 3 314 2 1 3 2 4 308...
Page 309: ...3 1 3 4 NPT 2 1 4 NPT 309...
Page 310: ...3 3 Polymetron 1 2 3 4 5 6 7 8 9 10 11 12 310...
Page 311: ...1 2 3 4 311...
Page 312: ...5 6 7 8 312...
Page 315: ...1 2 3 4 5 6 7 8 9 sc100 12 10 318 11 12 1 2 4 315...
Page 316: ...4 3 30 4 5 6 7 8 9 10 11 12 13 319 319 14 15 318 316...
Page 317: ...16 17 1 2 3 4 5 6 7 8 319 319 9 10 318 11 12 317...
Page 318: ...1 2 3 4 1 5 6 7 1 2 1 2 1 7 30 60 90 6 9 1 2 1 2 318...
Page 319: ...3 4 5 318 6 7 Modbus Modbus MSDS SDS 1 2 2 3 3 4 5 5 5 2 3 6 319...
Page 320: ...319 1 2 4 4 1089 1106 23 27 C 5 5 5 5 Sensor Days 320...
Page 322: ...7 100 C TEMP TOO LOW T 0 C 365 320 7 0 4 C 0 200 C 8 8 322...
Page 330: ...1 2 3 4 330 T rk e...
Page 331: ...5 6 7 8 T rk e 331...
Page 349: ...1 2 3 4 Slovensk jazyk 349...
Page 350: ...5 6 7 8 350 Slovensk jazyk...
Page 368: ...1 2 3 4 368 Slovenski...
Page 369: ...5 6 7 8 Slovenski 369...
Page 387: ...1 2 3 4 Hrvatski 387...
Page 388: ...5 6 7 8 388 Hrvatski...
Page 403: ...2 3 409 2 1 3 2 4 403...
Page 404: ...3 1 PVC 3 4 in NPT 2 1 4 in NPT 404...
Page 405: ...3 3 Polymetron 1 2 3 4 5 6 7 8 9 10 11 12 405...
Page 406: ...1 2 3 4 406...
Page 407: ...5 6 7 8 407...
Page 415: ...Borax 1 2 2 3 3 4 5 HCl 5 5 2 3 6 414 1 2 4 4 1089 1106 ohm 23 27 C 5 ohm 5 ohm 415...
Page 418: ...7 HOLD TC 0 4 C 0 200 C TC MENU 8 8 Event OK 8 Event K ADC ADC FLASH ERASE Flash 20 200 C 418...
Page 425: ...1 2 3 4 eesti keel 425...
Page 426: ...5 6 7 8 426 eesti keel...
Page 439: ...ESD 1 439...
Page 441: ...3 1 T PVC 3 4 NPT 2 T 1 4 NPT 441...
Page 442: ...3 3 Polymetron 1 2 3 Polymetron 3 4 5 6 7 8 9 10 11 12 442...
Page 443: ...1 2 3 4 443...
Page 444: ...5 6 7 8 444...
Page 454: ......
Page 455: ......