Option
Description
LOG SETUP
Sets the time interval for data storage in the data log
—5, 30 seconds, 1, 2, 5, 10, 15 (default), 30,
60 minutes.
RESET DEFAULTS
Sets the configuration menu to the default settings. All
user-defined settings are lost.
Adjust the T-factor for non-standard cable lengths
When the sensor cable is extended or shortened from the standard 6 m
(20 ft), the resistance of the cable changes. This change reduces the
accuracy of temperature measurements. To correct for this difference,
calculate a new T-factor.
Note: This procedure applies only to sensors with a PT1000 temperature element.
Sensors with a PT100 temperature element are less accurate.
1.
Measure the temperature of a solution with the sensor and with an
independent, reliable instrument such as a thermometer.
2.
Record the difference between the temperature measured from the
sensor and from the independent source (actual).
For example, if the actual temperature is 50 °C and the sensor
reading is 53 °C, the difference is 3 °C.
3.
Multiply this difference by 3.85 to get an adjustment value.
Example: 3 x 3.85 = 11.55.
4.
Calculate a new T-factor:
• Sensor temperature > actual—add the adjustment value to the T-
factor on the sensor cable
• Sensor temperature < actual—subtract the adjustment value from
the T-factor on the sensor cable
5.
Enter the new T-factor in the Configure, Temp Element menu.
Calibrate the sensor
About sensor calibration
During calibration, data is not sent to the datalog. Thus, the datalog can
have areas where the data is intermittent.
Zero calibration procedure
Use the zero calibration procedure to define the unique zero point of the
conductivity sensor. The zero point must be defined before the sensor is
calibrated for the first time with a reference solution or process sample.
1.
Remove the sensor from the process. Wipe the sensor with a clean
towel or use compressed air to make sure the sensor is clean and
dry.
2.
Push the
MENU
key and select Sensor Setup, [Select Sensor],
Calibrate.
3.
Push
ENTER
to select Zero Cal.
4.
If the passcode is enabled in the security menu for the controller,
enter the passcode.
5.
Select the option for the output signal during calibration:
Option
Description
Active
The instrument sends the current measured output value during
the calibration procedure.
Hold
The sensor output value is held at the current measured value
during the calibration procedure.
Transfer
A preset output value is sent during calibration. Refer to the
controller user manual to change the preset value.
6.
Hold the dry sensor in the air and push
ENTER
.
7.
Review the calibration result:
• Pass—the zero point is set.
• Fail—the value is outside of accepted limits. Make sure the sensor
is dry and repeat the zero calibration procedure. Make sure that
the cause is not the digital extension cable or a lot of electronic
noise.
8.
If the calibration passed, push
ENTER
to continue.
9.
For the sc100 controller, go to step
.
10.
If the option for operator ID is set to Yes in the Calibration Options
menu, enter an operator ID. Refer to
on page 14.
English
11
Summary of Contents for Polymetron Series
Page 8: ...1 2 3 4 8 English...
Page 9: ...5 6 7 8 English 9...
Page 26: ...1 2 3 4 26 Deutsch...
Page 27: ...5 6 7 8 Deutsch 27...
Page 46: ...1 2 3 4 46 Italiano...
Page 47: ...5 6 7 8 Italiano 47...
Page 64: ...1 2 3 4 64 Fran ais...
Page 65: ...5 6 7 8 Fran ais 65...
Page 81: ...Figura 3 Ejemplos de montaje 1 3 4 pulg NPT 2 1 4 pulg NPT Espa ol 81...
Page 83: ...1 2 3 4 Espa ol 83...
Page 84: ...5 6 7 8 84 Espa ol...
Page 103: ...1 2 3 4 Portugu s 103...
Page 104: ...5 6 7 8 104 Portugu s...
Page 121: ...1 2 3 4 e tina 121...
Page 122: ...5 6 7 8 122 e tina...
Page 139: ...1 2 3 4 Dansk 139...
Page 140: ...5 6 7 8 140 Dansk...
Page 158: ...1 2 3 4 158 Nederlands...
Page 159: ...5 6 7 8 Nederlands 159...
Page 178: ...1 2 3 4 178 Polski...
Page 179: ...5 6 7 8 Polski 179...
Page 196: ...1 2 3 4 196 Svenska...
Page 197: ...5 6 7 8 Svenska 197...
Page 214: ...1 2 3 4 214 Suomi...
Page 215: ...5 6 7 8 Suomi 215...
Page 230: ...2 3 236 2 1 3 2 4 230...
Page 231: ...3 1 Flow thru T PVC 3 4 NPT 2 Flow thru T 1 4 NPT 231...
Page 232: ...3 3 Polymetron No 1 2 3 4 5 6 7 8 9 10 11 12 232...
Page 233: ...1 2 3 4 233...
Page 234: ...5 6 7 8 234...
Page 247: ...2 ISO 7888 ASTM D5391 64 2 52 08394 A 2000 FDA EN 10204 3 1 B 0 4 m 64 2 52 08394 A 2011 247...
Page 254: ...1 2 3 4 254 Magyar...
Page 255: ...5 6 7 8 Magyar 255...
Page 273: ...1 2 3 4 Rom n 273...
Page 274: ...5 6 7 8 274 Rom n...
Page 292: ...1 2 3 4 292 lietuvi kalba...
Page 293: ...5 6 7 8 lietuvi kalba 293...
Page 308: ...2 3 314 2 1 3 2 4 308...
Page 309: ...3 1 3 4 NPT 2 1 4 NPT 309...
Page 310: ...3 3 Polymetron 1 2 3 4 5 6 7 8 9 10 11 12 310...
Page 311: ...1 2 3 4 311...
Page 312: ...5 6 7 8 312...
Page 315: ...1 2 3 4 5 6 7 8 9 sc100 12 10 318 11 12 1 2 4 315...
Page 316: ...4 3 30 4 5 6 7 8 9 10 11 12 13 319 319 14 15 318 316...
Page 317: ...16 17 1 2 3 4 5 6 7 8 319 319 9 10 318 11 12 317...
Page 318: ...1 2 3 4 1 5 6 7 1 2 1 2 1 7 30 60 90 6 9 1 2 1 2 318...
Page 319: ...3 4 5 318 6 7 Modbus Modbus MSDS SDS 1 2 2 3 3 4 5 5 5 2 3 6 319...
Page 320: ...319 1 2 4 4 1089 1106 23 27 C 5 5 5 5 Sensor Days 320...
Page 322: ...7 100 C TEMP TOO LOW T 0 C 365 320 7 0 4 C 0 200 C 8 8 322...
Page 330: ...1 2 3 4 330 T rk e...
Page 331: ...5 6 7 8 T rk e 331...
Page 349: ...1 2 3 4 Slovensk jazyk 349...
Page 350: ...5 6 7 8 350 Slovensk jazyk...
Page 368: ...1 2 3 4 368 Slovenski...
Page 369: ...5 6 7 8 Slovenski 369...
Page 387: ...1 2 3 4 Hrvatski 387...
Page 388: ...5 6 7 8 388 Hrvatski...
Page 403: ...2 3 409 2 1 3 2 4 403...
Page 404: ...3 1 PVC 3 4 in NPT 2 1 4 in NPT 404...
Page 405: ...3 3 Polymetron 1 2 3 4 5 6 7 8 9 10 11 12 405...
Page 406: ...1 2 3 4 406...
Page 407: ...5 6 7 8 407...
Page 415: ...Borax 1 2 2 3 3 4 5 HCl 5 5 2 3 6 414 1 2 4 4 1089 1106 ohm 23 27 C 5 ohm 5 ohm 415...
Page 418: ...7 HOLD TC 0 4 C 0 200 C TC MENU 8 8 Event OK 8 Event K ADC ADC FLASH ERASE Flash 20 200 C 418...
Page 425: ...1 2 3 4 eesti keel 425...
Page 426: ...5 6 7 8 426 eesti keel...
Page 439: ...ESD 1 439...
Page 441: ...3 1 T PVC 3 4 NPT 2 T 1 4 NPT 441...
Page 442: ...3 3 Polymetron 1 2 3 Polymetron 3 4 5 6 7 8 9 10 11 12 442...
Page 443: ...1 2 3 4 443...
Page 444: ...5 6 7 8 444...
Page 454: ......
Page 455: ......