Section 10 User configuration
User configuration
Default value User value
Date
ELECTRODE
NAME
Name
GLASS
Slope
Slope
+059.16
Uiso
Isothermal potential
+0000.0 mV
E
Reference potential
+0000.0 mV
C
Reference concentration
10E-8
OPTION
AUTOMAT. CAL YES/NO
Automatic calibration
NO
AUTO. CAL 1PT/2PTS
Type of automatic calibration
2 PTS
With sample / With rinse water
Calibration with sample or rinse water
With sample
PULSE PUMP YES/NO
Calibration with pulse pump
NO
Flow: XX.X ml/mn or l/puls
Flowrate of the pump
2 ml/minute
Cell Vol.: XXX.X ml
Volume of the cell
100 ml
STD ADDITION
Measurement with a known addition
NO
CLEANING YES/NO
Chemical cleaning
NO
SAMPLE COND YES/NO
Chemical conditioning
NO
MANUAL YES/NO
External sample measurement
NO
SAMPLE TIME
Sample injection time
10 seconds
HEATING YES/NO
Regulation in temperature
NO
Tc
Temperature value
25°C
DILUTION YES/NO
Dilution
NO
2CHANNELS YES/NO
Measurement of 2 different samples
NO
ADJUST Iout
OUT 4-20/0-20 mA
Output current
0-20 mA
Minimum value corresponding to 0 mA
0
Value 4 mA
795
Value 20 mA
3896
UNIT
ppb/ppm/USER
Unit
ppb
CLEANING: chemical cleaning
CLEAN. FREQ: XXX
Number of measurements between cleaning
cycles
5
SPRINKLER t: XXXs
Duration of the cleaning water sprinkler
5 seconds
REAG. INJ. t: XXXs
Duration of cleaning solution injection
5 seconds
RESIDENC. t: XXXs
Length of time the cleaning solution remains in
the reactor
5 seconds
53
Summary of Contents for POLYMETRON 8810
Page 1: ...DOC024 52 93057 POLYMETRON Model 8810 ISE Analyzer User Manual 05 2015 Edition 19 ...
Page 2: ......
Page 6: ...1 1 Dimensions Figure 1 Panel dimensions mm inches Specifications 4 ...
Page 7: ...Figure 2 Cabinet dimensions mm inches Specifications 5 ...
Page 8: ...Specifications 6 ...
Page 34: ...Menu overview 32 ...
Page 38: ...Maintenance and Troubleshooting 36 ...
Page 57: ......