Section 6
Menu overview
MAIN MENU
The main menu gives access to the following functions:
•
STOP
- Pressing
Enter
causes the measurement process to stop immediately
•
START
- Pressing
Enter
initiates the measurement process
•
STAND-BY
- Pressing
Enter
puts the analyzer in standby
•
PROGRAMMING
- Program the analyzer (
Command
mode only)
•
READING
- View parameters (
Operating
mode only)
In
PROGRAMMING
or
READING
mode the following options can be selected using the
scroll keys:
•
CALIBRATION
- Calibrate the analyzer
•
ANALOG-OUT
- Analog output parameters
•
ALARM
- Alarm limit parameters
•
SEQUENCE
- Measurement mode
•
SERVICE
- Check the analyzer and accessory operations
•
CLEANING
- Chemical cleaning parameters
•
SAMPLE COND.
- Sample conditioning parameters
•
CAL PARAMETER
- View the primary and last calibration details
•
TIMING
- Titration cycle parameters
•
PM XXXX
- System configuration (only available in
PROGRAMMING
mode and
requires a valid password)
Note: Some of the above options are only available if the relevant accessory has been installed.
6.1 System configuration (PM XXXX option)
Refer also to any
Installation Procedure
documents delivered with the analyzer for
application specific information.
1.
Select
PM XXXX
with the scrolling keys and press
Enter
. Using the
Select
key to
move from one digit to the next, enter the special code and press
Enter
.
Option
Description
ELECTRODE
Define the electrode parameters:
•
Name
•
Slope
•
(Uiso) Isothermal point potential
•
(E) Reference point potential adjusted to 25°C
•
(C) Reference point concentration adjusted to 25°C
The E and C parameters are used to calculate the standard potential
(
E
025
= E - S.logC
)
UNIT
Select the unit of concentration from the list available. Selecting
USER
allows you to define your own, and requires a 5 character description.
Adjust ADC
This option is reserved for Hach Lange service technicians.
CAL
TEMPERATURE
Two-point temperature calibration. Use a temperature simulator for
each of the two points. Wait until the measurement stabilizes before
accepting each calibration point.
Adjust Iout
Select the output type (0-20mA or 4-20mA). Place a multimeter in
series with the analog output and make any small adjustments to the
signal as required.
25
Summary of Contents for POLYMETRON 8810
Page 1: ...DOC024 52 93057 POLYMETRON Model 8810 ISE Analyzer User Manual 05 2015 Edition 19 ...
Page 2: ......
Page 6: ...1 1 Dimensions Figure 1 Panel dimensions mm inches Specifications 4 ...
Page 7: ...Figure 2 Cabinet dimensions mm inches Specifications 5 ...
Page 8: ...Specifications 6 ...
Page 34: ...Menu overview 32 ...
Page 38: ...Maintenance and Troubleshooting 36 ...
Page 57: ......