4. FILL BUFFER 2
is displayed. Pour the second calibration solution into the
measurement chamber and press
Enter
.
5. CONC 2 XX.XXEXX
is displayed. Enter the calibration solution concentration and
press
Enter
. The solution is then measured before the measurement chamber is then
cleaned.
6.2.2.2
Process calibration
1.
Select
PROCESS CAL
and press
Enter
.
2. Slope: XXX.XX
is displayed. Enter the slope value of the process solution. Press
Enter
to confirm.
3.
After measurement the concentration of the process solution is displayed
CONC
1/2 XX.XXEXX
. This value can be adjusted if necessary. Press
Enter
to confirm.
6.2.2.3
Automatic calibration
This is only available if the analyzer has this option installed. It allows for an automatic
calibration at pre-programmed intervals, using a calibration solution of known
concentration.
1.
Select
AUTOMAT. CAL
and press
Enter
.
2. FREQUENCY
is displayed. This defines the number of measurements that are made
before an automatic calibration takes place. Enter the number and confirm with
Enter
.
3.
If the
AUTOMAT CAL1 PT
option has been selected:
Option
Description
INJ. TIME
Enter the injection time of the calibration solution in seconds and confirm with
Enter
.
CONC
Enter the concentration of the calibration solution and press
Enter
.
IMMEDIATE
Select
YES
or
NO
followed by
Enter
. If
YES
is selected an automatic
calibration will start immediately. No more input is required.
4.
If the
AUTOMAT CAL 2PTS
option has been selected, select
CAL PUMP YES
to
start a pump calibration (pulse or micro system),
CAL PUMP NO
if no calibration is
required or
Use Default Parameters
to use the parameters defined in
PMXXX-
OPTION-AUTOMAT.CAL
and press
Enter
.
5.
If the
CAL PUMP YES
option has been selected:
Option
Description
INJ.TIME or
INJECT
Enter the injection time (micro system pump) in seconds or the number
of pulses (pulse pump) and press
Enter
.
ADD C
Enter the concentration of the addition and press
Enter
. The pump
calibration starts.
Fill C. PUMP SOL
Press
Enter
for the introduction of the solution with known
concentration. The concentration of the solution in the measurement
chamber is then measured, followed by an injection of the calibration
solution, followed by another measurement of the solution in the
measurement chamber.
Menu overview
27
Summary of Contents for POLYMETRON 8810
Page 1: ...DOC024 52 93057 POLYMETRON Model 8810 ISE Analyzer User Manual 05 2015 Edition 19 ...
Page 2: ......
Page 6: ...1 1 Dimensions Figure 1 Panel dimensions mm inches Specifications 4 ...
Page 7: ...Figure 2 Cabinet dimensions mm inches Specifications 5 ...
Page 8: ...Specifications 6 ...
Page 34: ...Menu overview 32 ...
Page 38: ...Maintenance and Troubleshooting 36 ...
Page 57: ......