Option
Description
C. Ini
Adjust the concentration of the solution before the addition if necessary.
C. End
Adjust the concentration of the solution after the addition if necessary.
6.
If the
CAL PUMP NO
or the
Use Default Parameters
option has been selected:
Option
Description
ADD C
Enter the concentration of the addition and press
Enter
. The pump
calibration starts.
INJ.TIME1 or
INJECT1
Enter the injection time (micro system pump) in seconds or the number
of pulses (pulse pump) for the first calibration solution and press
Enter
.
INJ.TIME2 or
INJECT2
Enter the injection time (micro system pump) in seconds or the number
of pulses (pulse pump) for the second calibration solution and press
Enter
.
IMMEDIATE
Select
YES
or
NO
followed by
Enter
. If
YES
is selected an automatic
calibration will start immediately. No more input is required.
6.2.3 Calibration results
The following tolerances are defined for calibrations:
•
Zone 1: The new calibration parameters are between 70% and 130% of the current
calibration parameters
•
Zone 2: The new calibration parameters are between 50% and 70% or 130% and
150% of the current calibration parameters
•
Zone 3: The new calibration parameters are between <50% and >150% of the current
calibration parameters
If the new calibration parameters are in
Zone 1
, the calibration is considered successful
and the new parameters are used for all future measurements.
If the new calibration parameters are in
Zone 2
, it is necessary to
CONFIRM
the new
parameters. If
YES
is selected the new parameters are used for all future measurements.
If
NO
is selected the new parameters are rejected and the previous values are used for
all future measurements.
If the new calibration parameters are in
Zone 3
, the message
ERROR CALIBRAT.
is
displayed. Press
Enter
to continue then select
RETRY
to repeat the calibration or
ABORT
to abort the calibration.
These messages only appear for a manual calibration, a process calibration, or if
IMMEDIATE YES
was selected for an automatic calibration.
In a standard automatic calibration, or when
IMMEDIATE NO
was selected, the analyzer
will not ask for confirmation. If the calibration parameters are within
Zone 1
the calibration
is considered successful and the new parameters are used for all future measurements. If
the calibration parameters are within
Zone 2
an error message
SLOPE CALIB
appears
on the display and the new calibration parameters are rejected. f the calibration
parameters are within
Zone 3
an error message
ERROR CALIB
appears on the display
and the new calibration parameters are rejected.
6.3 Analog outputs (ANALOG-OUT option)
The analyser has two analog outputs which are allocated according to the application:
•
One connected to the concentration
•
The other connected to the potential electrode line
For logarithmic and bilinear outputs, the minimum range is 1/100 of the maximum range.
The minimum range cannot be adjusted.
Menu overview
28
Summary of Contents for POLYMETRON 8810
Page 1: ...DOC024 52 93057 POLYMETRON Model 8810 ISE Analyzer User Manual 05 2015 Edition 19 ...
Page 2: ......
Page 6: ...1 1 Dimensions Figure 1 Panel dimensions mm inches Specifications 4 ...
Page 7: ...Figure 2 Cabinet dimensions mm inches Specifications 5 ...
Page 8: ...Specifications 6 ...
Page 34: ...Menu overview 32 ...
Page 38: ...Maintenance and Troubleshooting 36 ...
Page 57: ......