J4, J5, J6 and J7 Terminal number
Color
1
Not used
2
Not used
3
Brown
4
White
4.5.3.5
RS232 connection
Refer also to
Figure 8
on page 18 and
Table 4
on page 19. Wire the connection to J10 as
follows:
J10 Terminal number
DB9 Plug
DB25 Plug
1 (TXD)
RXD: 2
RXD: 2
2 (RXD
TXD: 3
TXD: 3
3 (GND)
COM: 5
COM: 7
The 8810 configuration is:
•
Speed: 9600 baud
•
Data: 8 bits
•
Stop bit: 2
•
Parity: none
On startup the analyzer sends the name and software version:
•
ANALYZER 8810 : 00 : 00 : 00 :
•
ISE X.XX : 00 : 00 : 00 :
In measurement mode, the analyzer sends the following data:
•
Measure HH : MM : SS
•
M1 XXXXEXX XX.X°C XXX
Where:
•
MX = M1 for channel 1 or M2 for channel 2
•
XXXXEXX = concentration measurement value
•
XX.X°C = temperature
•
XXX = value of the potential
Where appropriate, the analyzer also sends the following messages with a time stamp
(HH : MM : SS):
ACTIONS
Cleaning
Chemical cleaning
Stopped
Analyzer stopped
Fix-time
Time between two cycles
Standby
Analyzer in standby
AUTO SYS CAL
Automatic calibration
PROCESS SYS CAL
Process calibration
MANUAL SYS CAL
Manual calibration
Installation
21
Summary of Contents for POLYMETRON 8810
Page 1: ...DOC024 52 93057 POLYMETRON Model 8810 ISE Analyzer User Manual 05 2015 Edition 19 ...
Page 2: ......
Page 6: ...1 1 Dimensions Figure 1 Panel dimensions mm inches Specifications 4 ...
Page 7: ...Figure 2 Cabinet dimensions mm inches Specifications 5 ...
Page 8: ...Specifications 6 ...
Page 34: ...Menu overview 32 ...
Page 38: ...Maintenance and Troubleshooting 36 ...
Page 57: ......