12
Specifications
Data storage
Rolling buffer or store once mode for up to 1000 measurements and 1000 operator actions
Holds calibration records for the last 10 calibrations
User interface
Touch-screen panel
Displays concentration, trend graph, diagnostics,
alarm status, historical data.
Password protection
Five levels of authorised access to configuration and
data management
Sensor
Sample temperature
Measurement from -5 to 50°C (23 to 122°F)
Sensor resistant to temperature from -5 to 100°C (23 to 212°F)
Sample pressure
1 to 20 bar absolute (14.5 to 290 psia)
Sample flow rate requirement
50 to 300 mL/min
Sample types
K1100 sensor
Water only
M1100 sensor (low range)
Water and beer
M1100 sensor (high range)
Water, beer, wine, wort and carbonated drinks
CIP
K1100 sensor
Not designed for use in CIP processes
M1100 sensor
Resistant to all common CIP methods
Measurement range
Low range sensors: 0 to 2,000 ppb (dissolved. Indicatives values until 5000 ppb)
High range sensors: 0 to 40 ppm (dissolved)
Repeatability
Low range sensors: ± 0.4 ppb or 1%, whichever is the greater
High range sensors: ± 0.015 ppm or ± 2% whichever is the greater
Reproducibility
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Accuracy
Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater
Limit of detection (LOD)
Low range sensors: 0.6 ppb
High range sensors: 0.015 ppm
Response time (90%)
Low range sensors: < 10 seconds in gas phase; < 30 seconds in liquid phase
High range sensors: < 10 seconds in gas phase; < 50 seconds in liquid phase
Warm-up time
None
Display resolution
0.1 ppb
Calibration
Low range sensors: Single point calibration (zero)
High range sensors: Two at cap replacement (zero and air), one during use (air)
Calibration sample
Low range sensors: Standard 99.999% N
2
(quality 50) or equivalent oxygen free gas
High range sensors: Standard 99.999% N
2
(quality 30) or equivalent oxygen free gas, air
Accessories
Accessories
Active spots, spare sensors, tool kit, portable calibration setup, etc.
Ask your local Hach representative for more details on all available spare parts and
accessories.
Installation
In process sensor installation
devices
ORBISPHERE insertion/extraction valve for installation on Varinline® access units
ORBISPHERE 28 mm sensor weld-on socket
PG 13.5 stationary housing for installation on Varinline® access units
Note:
This device is only available for the M1100 sensor
28 mm stationary housing for installation on Varinline® access units
Ambient temperature
-5 to 50 °C (23 to 122 °F)
Humidity
0 to 95% non-condensing relative humidity
Power supply
Universal 85-264 VAC @ 50/60 Hz, 25 VA
10-36 VDC, 25 W
Weights
Wall and pipe mount
(H x D x W)
236.5 x 160 x 250 mm - weight 4.25 kg
9.31 x 6.30 x 9.84 in. - weight 8.82 lbs
Summary of Contents for ORBISPHERE K-M1100
Page 5: ...4 Table of Contents ...
Page 19: ...18 Specifications ...
Page 39: ...38 Installation ...
Page 45: ...44 User Interface ...
Page 48: ...47 Section 6 View Menu Figure 33 View menu ...
Page 51: ...50 View Menu ...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu ...
Page 59: ...58 Measurement Menu ...
Page 65: ...64 Calibration Menu ...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 39 Inputs Outputs menu ...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 40 Analog outputs menu ...
Page 75: ...74 Inputs Outputs Menu ...
Page 91: ...90 Communication Menu ...
Page 94: ...93 Section 12 Products Menu Figure 43 Products menu ...
Page 97: ...96 Global Configuration Menu ...
Page 98: ...97 Section 14 Services Menu Figure 45 Services menu part 1 ...
Page 99: ...98 Services Menu Figure 46 Services menu part 2 ...
Page 107: ...106 Maintenance and Troubleshooting ...
Page 113: ...112 Accessories and Spare Parts ...