73
Inputs/Outputs Menu
Tri-linear
Range
Measurement M
Resolution R
4-20 mA
1: AOL > I > 4
M=MLL+(ML-MLL)
*
(I-4)/(AOL-4)
R=(ML-MLL)
*
20/((AOL-4)
*
1010)
2: AOH > I > AOL
M=ML+(MH-ML)
*
(I-AOL)/ (AOH-AOL)
R=(MH-ML)
*
20/((AOH-AOL)
*
1010)
3: 20 > I > AOH
M=MH+(MHH-MH)
*
(I-AOH) / (20-AOH) R=(MHH-MH)
*
20/((20-AOH)
*
1010)
0-20 mA
1: AOL > I > 0
M=MLL+(ML-MLL)
*
I/AOL
R=(ML-MLL)
*
20/(AOL
*
1010)
2: AOH > I > AOL
M=ML+(MH-ML)
*
(I-AOL)/ (AOH-AOL)
R=(MH-ML)
*
20/((AOH-AOL)
*
1010)
3: 20 > I > AOH
M=MH+(MHH-MH)
*
(I-AOH)/(20-AOH)
R=(MHH-MH)
*
20/((20-AOH)
*
1010)
Analog output characteristics (continued)
None
This is the default value.
Setting the analog output to “None” means that the output value will always
be zero and importantly ensures that no current is emitted on that output, so
reducing power consumption as well as reducing heat within the instrument.
Summary of Contents for ORBISPHERE K-M1100
Page 5: ...4 Table of Contents ...
Page 19: ...18 Specifications ...
Page 39: ...38 Installation ...
Page 45: ...44 User Interface ...
Page 48: ...47 Section 6 View Menu Figure 33 View menu ...
Page 51: ...50 View Menu ...
Page 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu ...
Page 59: ...58 Measurement Menu ...
Page 65: ...64 Calibration Menu ...
Page 66: ...65 Section 9 Inputs Outputs Menu Figure 39 Inputs Outputs menu ...
Page 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 40 Analog outputs menu ...
Page 75: ...74 Inputs Outputs Menu ...
Page 91: ...90 Communication Menu ...
Page 94: ...93 Section 12 Products Menu Figure 43 Products menu ...
Page 97: ...96 Global Configuration Menu ...
Page 98: ...97 Section 14 Services Menu Figure 45 Services menu part 1 ...
Page 99: ...98 Services Menu Figure 46 Services menu part 2 ...
Page 107: ...106 Maintenance and Troubleshooting ...
Page 113: ...112 Accessories and Spare Parts ...