Communication Menu
9.1.2 Example of use
In this example we use:
•
One PC with a RS232 port.
•
One "RS-485<->RS232 converter"
Procedure:
1.
Connect both RS-485 wires of the instrument to the "RS-485<->RS232 converter".
2.
Connect the "RS-485<->RS232 converter" to the PC RS232 port using a standard cable
(RS232 DB9 straight cable).
On the PC:
1.
Run "Hyperterminal" on the PC.
2.
Configure the PC COM port used (e.g. COM2). Menu "File/Properties/Configure".
3.
Configure the parameters "Baud rate", "Parity", "Nb of stop bits", "Byte size" (Menu
"File/Properties/Configure"). Use the same parameters for the instrument and the PC.
4.
Configure the "Font = Courier 10" (Menu "View/Font").
5.
Connect "Hyperterminal" (Menu "Call/Call").
6.
Save the data received in the file of your choice (Menu "Transfer/Capture Text/Start").
On the instrument:
1.
Use the menu "Communication/RS-485 Simple/Send files" and the button "All files".
When the transfer is finished, close the file with "Hyperterminal" (Menu "Transfer/Capture
Text/Stop"). Now, all the reports are saved in a text file on your PC.
Configuration report example
The "Configuration" below is given for a one channel instrument.
INSTRUMENT CONFIGURATION
Measurement mode . . . . . . . . . Continuous mode
Pressure unit. . . . . . . . . . . . . . . [bar]
Temperature unit . . . . . . . . . . . . [°C]
Storage mode . . . . . . . . . . . . . . Rolling buffer
Storage RAM time. . . . . . . . . . . 10 [s]
Storage FLASH time . . . . . . . . . 3600 [s]
Channel 1
Membrane. . . . . . . . . . . . . . . . . 2956A
Medium . . . . . . . . . . . . . . . . . . . Liquid
Gas unit. . . . . . . . . . . . . . . . . . . mbar
Liquid . . . . . . . . . . . . . . . . . . . . Water
Resolution displayed. . . . . . . . . 3
Thermal cut off . . . . . . . . . . . . . Enabled60.0 [°C]
Alarms
Low Low . . . . . . . . . . . . . . . . . . Disabled0.000000 [mbar]
Low . . . . . . . . . . . . . . . . . . . . . . Disabled0.100000 [mbar]
High. . . . . . . . . . . . . . . . . . . . . . Enabled5.000000 [mbar]
High High . . . . . . . . . . . . . . . . . Disabled1.100000 [mbar]
Hysteresis . . . . . . . . . . . . . . . . . 1 [%]
Delay. . . . . . . . . . . . . . . . . . . . . 0 [s]
Filter State . . . . . . . . . . . . . . . . . Disabled
Type . . . . . . . . . . . . . . . . . . . . . Median
Depth . . . . . . . . . . . . . . . . . . . . 5
Central depth . . . . . . . . . . . . . . 3
Interferences
CO
2
or H
2
S . . . . . . . . . . . . . . . . All disabled
Chlorinity/Salinity . . . . . . . . . . . All disabled19.000000 [g/l]
H
2
. . . . . . . . . . . . . . . . . . . . . . . Disabled0.100000 [bar]
Summary of Contents for ORBISPHERE 410
Page 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Page 39: ...38 User Interface...
Page 40: ...39 Section 5 View Menu Figure 25 View menu...
Page 43: ...42 View Menu...
Page 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Page 49: ...48 Measurement Menu...
Page 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Page 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Page 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Page 65: ...64 Inputs Outputs Menu...
Page 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Page 85: ...84 Global Configuration Menu...
Page 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Page 87: ...86 Services Menu Figure 39 Services menu Part 2...
Page 91: ...90 Services Menu...
Page 95: ...94 Part Lists...