27
Installation
3.6.2 Power supply connection (high voltage instruments)
High voltage instruments (100-240 VAC) have a 4-pin male connector pre-wired internally with a
male BINDER connector ready for mains connection. A compatible female connector is supplied
with the instrument.
If this female connector was supplied with a mains power plug already pre-attached (cable part
numbers 33031, 33032, 33033 and 33034) then the female connector can be plugged directly
into the instrument power connector. The two connectors are grooved to avoid an incorrect
fitting. Tighten the female connector to the instrument power connector finger-tight.
If no power cable was ordered with the equipment, a mains power plug must be connected to
the supplied female connector as described in the following procedure.
User-supplied power cable specifications:
•
3-wire (live, neutral and earth)
•
cable
ø
7mm;
9.5mm
•
wire selection
mm
2
, AWG18;
2.5mm
2
, AWG14
Prepare the user-supplied power cable as follows:
Strip off 23 mm (0.9 ins.) of shielding from the power cable. Cut back the live and neutral wires
to 15 mm (0.6 ins.) in length but leave the earth wire as is. Then strip off a small amount of
external insulation from the three wires as required:
CAUTION
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this
section of the manual.
WARNING
Potential Electrocution Hazard. Always disconnect power to the instrument when
making electrical connections.
Summary of Contents for ORBISPHERE 410
Page 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Page 39: ...38 User Interface...
Page 40: ...39 Section 5 View Menu Figure 25 View menu...
Page 43: ...42 View Menu...
Page 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Page 49: ...48 Measurement Menu...
Page 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Page 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Page 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Page 65: ...64 Inputs Outputs Menu...
Page 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Page 85: ...84 Global Configuration Menu...
Page 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Page 87: ...86 Services Menu Figure 39 Services menu Part 2...
Page 91: ...90 Services Menu...
Page 95: ...94 Part Lists...