Inputs/Outputs Menu
Analog outputs (continued)
Calibration of the analog
output
The calibration of the analog output is aimed at aligning the internally
calculated current to the real current output. This was performed at factory,
but could become necessary again because of electronic tolerances. A
precision amperometer (or voltmeter for the voltage versions) connected at
the corresponding analog output connection point is required. Refer to
Measurement board on page 30
.
•
Select the analog output channel to calibrate and press the
Start
button.
•
Measure with the amperometer the current value for point 1. It should be
below 4 mA
•
Edit point 1 and enter the same value as read on the amperometer, then
press the
Continue
button.
•
Measure with the amperometer the current value for point 2. It should be
above 20 mA.
•
Edit point 2 and enter the same value as read on the amperometer,
before pressing the
Continue
button.
Calibration of the selected analog output channel is completed.
Direct test
Test to check the calibration of the analog outputs. A precision amperometer
connected at the analog output connection point is required.
•
Select a value (4, 12, 20 mA available) for each channel and compare
this value (+ 0.02 mA) with what the amperometer shows.
A calibration is required if the value on the amperometer differs from the
current se 0.02 mA).
Note:
It is possible to test one analog output without interfering with the
others. During the test, the other analog outputs will continue to indicate the
measurement.
Characteristics test
This is a test for the correct operation of the peripherals connected to each
analog output, by verifying that the PLC computes the correct value.
The analog output will send the current corresponding to the value entered
in the text boxes.
•
Type in a test value for each analog output, and check for the related
action on the peripheral.
Summary of Contents for ORBISPHERE 410
Page 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Page 39: ...38 User Interface...
Page 40: ...39 Section 5 View Menu Figure 25 View menu...
Page 43: ...42 View Menu...
Page 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Page 49: ...48 Measurement Menu...
Page 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Page 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Page 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Page 65: ...64 Inputs Outputs Menu...
Page 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Page 85: ...84 Global Configuration Menu...
Page 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Page 87: ...86 Services Menu Figure 39 Services menu Part 2...
Page 91: ...90 Services Menu...
Page 95: ...94 Part Lists...