67
Communication Menu
9.1.1 Data available
All individual data are separated by at least one tabulation character (ASCII code=0x09).
For the cyclic measurements, the data format is detailed. For the files, only one example for
each file is given to explain the data format.
Cyclic measurements
1.
If the option "Mode = standard" is chosen, the following message is sent:
with:
\t ................................................the ASCII tab character: code=0x09
\r ................................................the ASCII Carriage Return character: code=0x0D
\n ..............................................the ASCII Line Feed character: code=0x0A
CHn ..........................................the 2 ASCII characters "CH" + the channel no.
Gas ............................................the gas concentration.
Gas Unit ....................................the gas unit.
Temperature ..............................the temperature.
Temperature Unit.......................the temperature unit.
Barometric Pressure..................the barometric pressure.
Barometric Pressure Unit ..........the barometric pressure unit.
Event .........................................the event bit mask in hexadecimal format.
The values are not described here. Refer to
List of events and alarms on page 92
.
•
Example of one measurement:
CH1 697.176 mbar 20.1 °C 0.982 bar C00
2.
If the option "Mode = expert" is chosen, the following message is sent:
with:
Current ......................................The current measured in [microA].
Partial pressure .........................The partial pressure in [bar].
External Pressure .....................The external pressure.
External Pressure Unit...............The external pressure unit.
Time ..........................................The time of the measurement. Format "hh:mm:ss."
Index..........................................This is the index of the last measurement.
This number starts at 0 at power up of the program.
•
Example of one measurement:
CH1 697.173 mbar 20.1 °C 0.982 bar C00 80.056229 µA 0.697
bar 1.000 bar 12:59:42 5923
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t Barometric Pressure\t Barometric Pressure Unit\t\ Event\t\r\n
CHn\t Gas\t Gas Unit\t Temperature\t Temperature Unit\t BarometricPressure\t Barometric Pressure Unit\t
Event\t Current\t µA\t Partial pressure\t bar\t External Pressure\t External Pressure Unit\t Time\t Index\r\n
Summary of Contents for ORBISPHERE 410
Page 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Page 39: ...38 User Interface...
Page 40: ...39 Section 5 View Menu Figure 25 View menu...
Page 43: ...42 View Menu...
Page 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Page 49: ...48 Measurement Menu...
Page 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Page 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Page 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Page 65: ...64 Inputs Outputs Menu...
Page 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Page 85: ...84 Global Configuration Menu...
Page 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Page 87: ...86 Services Menu Figure 39 Services menu Part 2...
Page 91: ...90 Services Menu...
Page 95: ...94 Part Lists...