74
Communication Menu
Activate sensor command - output
Activate sensor command - input
Set channel hold status- output
Set channel hold status - input
Name
Value
Comment
OCT
1-2
OCI
2
Command ID
OCD1
0
Channel number:
0 = Channel 1
OCD2
0-1
Sensor activation:
0 = Deactivate the EC sensor
1 = Activate the EC sensor
OCD3
Not used
OCD4
Not
used
Name
Value
Comment
ICT
1-2
ICS
0-3
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no)
3 = Execution failure
ICD1
Not used
ICD2
Not used
ICD3
Not used
ICD4
Not
used
Name
Value
Comment
OCT
1-2
OCI
3
Command ID
OCD1
0
Channel number:
0 = Channel 1
OCD2
0-1
Status:
0 = Channel enabled
1 = Channel on hold
OCD3
Not used
OCD4
Not
used
Name
Value
Comment
ICT
1-2
ICS
0-3
0 = OK
1 = Unknown command ID
2 = Invalid parameter (e.g. invalid channel no)
3 = Execution failure
ICD1
Not used
ICD2
Not used
ICD3
Not used
ICD4
Not
used
Summary of Contents for ORBISPHERE 410
Page 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Page 39: ...38 User Interface...
Page 40: ...39 Section 5 View Menu Figure 25 View menu...
Page 43: ...42 View Menu...
Page 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Page 49: ...48 Measurement Menu...
Page 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Page 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Page 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Page 65: ...64 Inputs Outputs Menu...
Page 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Page 85: ...84 Global Configuration Menu...
Page 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Page 87: ...86 Services Menu Figure 39 Services menu Part 2...
Page 91: ...90 Services Menu...
Page 95: ...94 Part Lists...