Specifications
2.2 Hardware description
The instrument hardware is made of one main board, and one measurement board for the
measurement channel (= the sensor).
The main board includes the controls for power, display, the touch screen, the barometric
sensor, the alarms, and communication ports. The measurement board performs measurements
and executes commands from the main board. It holds the "Analog output" and "Relays" that
send information to external systems.
A hardware watchdog is activated at program start up, to check that the system is not frozen
(i.e. infinite loop, system crash, etc.). If the watchdog is not refreshed by the software every
minute, the measurement display, the relays and the analog output are frozen up to 2 minutes.
Then the reset shuts down the instrument for 10 seconds and the start-up procedure is
performed. At the same time all the hardware (sensor, measurement board) are reset.
2.3 Model identification system
The analyzer identification number and the instrument serial number are located on the label on
the back panel, and can be found on order confirmation and invoice papers.
System alarm relay on the main
board
One "instrument system alarm" relay per instrument
1A-30 VAC or 0.5A-50 VDC on a resistance load
Normally closed [NC] (NO relay also available) when instrument is turned on. Opens when
a system alarm is detected, and when it does not receive any signal.
WARNING
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC RMS.
COMMUNICATION
Options
RS-485 or PROFIBUS-DP (optional); USB host ; Ethernet 10/100 Base-T
GENERAL
Thermal cut off
Prevents ageing of sensors when exposed to high temperatures
SIZE AND WEIGHT
Wall and pipe mount
(H x D x W)
236.5 x 160 x 250 mm - weight 4.25 kg
9.31 x 6.30 x 9.84 in. - weight 8.82 lbs
Panel mount: Face (housing)
(H x D x W)
156 (123) x 250 x 220 (214) mm - weight 3.35 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) in. - weight 6.62 lbs
A Oxygen EC sensor
C Ozone EC sensor
D Carbon dioxide TC sensor
W Wall
P Panel
1
100-240 VAC
2
10-30 VDC
C 0/4-20 mA
V 0-5 V
0
RS-485
1
PROFIBUS-DP / RS-485
0
Standard
K Korean language
4
1
0
/
/
0
0
0
Summary of Contents for ORBISPHERE 410
Page 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Page 39: ...38 User Interface...
Page 40: ...39 Section 5 View Menu Figure 25 View menu...
Page 43: ...42 View Menu...
Page 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Page 49: ...48 Measurement Menu...
Page 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Page 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Page 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Page 65: ...64 Inputs Outputs Menu...
Page 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Page 85: ...84 Global Configuration Menu...
Page 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Page 87: ...86 Services Menu Figure 39 Services menu Part 2...
Page 91: ...90 Services Menu...
Page 95: ...94 Part Lists...