73
Advanced Operations
•
Should individual program sequences be repeated?
The elements of a measuring sequence, such as zero and sample
measurements and the timer(s) (reaction times, waiting times, etc.)
are individually defined.
6.1.2.2 Enter a new element of a measuring sequence
Important Note:
Each component of the measurement process
must
be entered in the order in which it will be completed.
1.
Highlight the
Measurement Process
line in the data overview
and press
Edit
.
2.
Press
Edit
again and then
New
.
Content and definition of the keys
[Z]
key /
Zeroing
1.
Press the
[Z]
key to program a zero measurement. Confirm
with
OK
.
2.
Press
New
and then
Zeroing...
and use the alphanumeric
keypad to enter the wavelength at which the zero measurement
is to be completed. Press
OK
and confirm the input by pressing
OK
again.
3.
If zero measurements are to be carried out at a number of
wavelengths, repeat the above two steps for each wavelength.
Note:
The entered measurement sequence is displayed.