3.4  Cabling of the CM02 / CM02BP and the RM02  

 (diagram)

WARNING!

• 

Only use the supplied USB 3.0 cable (

REF 1022373

) for the connec‑

tion to the PC.

• 

Only use medically approved PCs or operate via a medically ap‑

proved isolating transformer.

• 

Auxiliary units on the PC (e.g. printer, monitor) must be operated 

through an isolating transformer. 

• 

Ethernet may only be used through a galvanic isolation in accor‑

dance with EN 60601‑1.

• 

The power supply unit's mains connector must be accessible in order 

to allow for disconnection from the electric mains at any time!

24. Power grid

25. Med. approved isolating transformer

26. Instrument table (IT)

27. Med. approved power supply

28. Printer

29. Screen

30. Personal computer

31. Local network

32. Galvanic isolation (EN 60601‑1) 

33. Headrest 

34. Cable headrest / LED illumination / 

RM02

35. Release module RM02

36. LED illumination

37. Camera module CM02 / CM02BP

24

25

26

28

29

30

33

35

36

37

31

32

34

27

NOTE!

To ensure that the system works correctly, Haag‑Streit recommends 

not using laptops and using a high‑quality desktop computer instead.

38. Fastening screw

39. Two‑pole connection plug

If the middle LED lights up red during 

operation, the two‑pole connection 

plug (39) is connected incorrectly. 

• 

Disconnect the device from the pow‑

er grid.

• 

Remove the cover on the upper part 

of the illumination facility by loosen‑

ing the fastening screw (38).

180°

38

39

1

 

• 

Turn the two‑pole connection plug (39) 180°. 

• 

Fix the cover on the upper part of the illumination facility with the fastening screw 

(38).

• 

Connect the device to the power grid again.

WARNING!

• 

Keep magnet-sensitive storage media (e.g. credit cards) away from 

the magnets on the release module RM02.

• 

Only external medical power supplies approved by Haag‑Streit that 

fulfill EN 60601-1 may be used.

3.4.1  Step-by-step cabling

• 

Place the release module RM02 over the slit lamp's cross slide. Four magnets are 

used for fixing.

NOTE!

• 

No external USB devices may be connected to USB ports (58) and 

(60). 

• 

With BQ 900 slit lamps with a date of manufacture before 1998, the 

cover plate is fixed on the crosstree carriage with screws. The two 

screws at the back must be removed before the RM02 is mounted.

• 

Insert the camera cable connector plug (49) in the socket (58).

• 

Press the camera cable into the cable holders (48).

• 

Computer cable USB 3.0 (46), power supply cable (45) and cable headrest / LED 

illumination / RM02 (44) must be fed through in the braided sleeving (43).

• 

Insert the connector plug of the power supply cable (45) in the socket (59).

01-IFU_IM600-7220506-04050_eng.indd   8

11.06.2019   16:03:02

Summary of Contents for IM 600

Page 1: ...A HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 DOC No 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Imaging Module IM 600 Accessory for slit lamps BQ 900 and BP 900 5 edi...

Page 2: ...urpose of use This device is an accessory for Haag Streit slit lamps BQ 900 or BP 900 which can be used to produce digital photographs and videos for documentation of the eye Contraindication There ar...

Page 3: ...igger 10 5 4 White balance 10 5 4 1 Slit lamp preparation 10 5 4 2 Conducting a white balance 10 5 5 Software Help menu error messages 11 5 6 LED display illumination head 11 5 7 LED display power sup...

Page 4: ...hPa 95 Use Temperature Air pressure Relative humidity from from from 10 C 800 hPa 30 to to to 35 C 1060 hPa 90 Air pressure Relative humidity from from 500 hPa 10 to to 1060 hPa 95 Storage Temperature...

Page 5: ...xpressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller This product is covered b...

Page 6: ...umination see separate instructions for use 3 Fiber optic line for periphery or background illumination 4 Headrest see separate instructions for use 5 Rail cover 2 4 Camera module CM02 CM02BP 6 Camera...

Page 7: ...at the corner opposite the black surface Position the sticker against the right roller rail 18 and the gliding plate 19 Press firmly on the white black surface press away any air bubbles Carefully te...

Page 8: ...connection plug If the middle LED lights up red during operation the two pole connection plug 39 is connected incorrectly Disconnect the device from the pow er grid Remove the cover on the upper part...

Page 9: ...se module RM02 51 RM02 operational control LED 52 On Off key 53 Rotating knob periphery or back ground illumination 54 Rotating knob slit illumination 55 Selector key A 56 Selector key B 57 Release ke...

Page 10: ...ease key 57 on the release module RM02 again the image is saved 5 4 White balance The Haag Streit IM 600 is optimized for maximum image quality with the Haag St reit slit lamp BQ 900 and BP 900 The im...

Page 11: ...illumination Polarity Slit lamp illumination a b c Operating status Standby mode Green short flashes Green short flashes Normal operation Slit and periphery or background illumi nation on Green Green...

Page 12: ...ached Red flashing Red flashing E4 No communication between power supply and illumination Contact your Haag Streit representative Red flashing 2 Red flashing 2 E6 General error Send PS LED to the appr...

Page 13: ...M if third party applications such as patient ad ministration software are to be used alongside EyeSuite Use 2 RAM module Hard disk At least 500 GB NTFS data system Graphics Graphics card with at leas...

Page 14: ...nst liquids A Appendix A 1 Accessories consumables spare parts upgrade Components REF Set white balance greycards 1021485 Power supply ICCNEXERGY 1022106 Computer cable USB 3 0 1022373 Release module...

Page 15: ...be installed in accordance with the EMC instruc tions contained in this accompanying document The operation of other lines or equipment than those listed may lead to higher emissions or may reduce th...

Page 16: ...t of a typical commercial or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 1 kV for symmetrical voltages Mains power quality should be that of a typical commercial or hospital...

Page 17: ...aration distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference...

Page 18: ...output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 7 80 MHz 800 MHz D 1 2 800 MHz 2 5 GHz D 2 3 0 01 0 1 1 10 100 For transmitters with a nominal outpu...

Page 19: ...EUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 01 IFU_IM600 7220506 04050_eng indd 19 11 06 2019 16 0...

Page 20: ...u have any further questions please contact your Haag Streit representative at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerla...

Reviews: