5.1 Camera Module CM02
Set the switch (rotating knob) beam splitter to the
camera symbol
•
70% of the light goes to the camera and
•
30% to the examiner
Switch (rotating knob) beam splitter to the top
•
100% of the light goes to the examiner (applies to both
beam paths)
5.2 Field of view
WARNING!
The images and videos should only be used for documentation purpo-
ses. Only the image in the eyepiece may be used for diagnosis.
Field of view of the object, see table
Circle:
The field of view of the object observed through the microscope's
eyepiece.
Rectangle: Surface area of the sensor:
12.5 ×
7.2 × 5.4
6.3 ×
10 ×
16 ×
25 ×
40 ×
ø 33*
ø 20
ø 12.7
ø 8
ø 5.1*
25.3 × 33.8*
16.0 × 21.3
10.1 × 13.5
6.4 × 8.5
4.1 × 5.4*
Object image section in the
eyepiece (mm)
Object image section in the
camera with sensor size
(mm)
* Not for slit lamp BP 900
5.3 History trigger
•
Press the release key (57) on the release module RM02
•
Select the desired image via selector key (55) or (56)
•
Press the release key (57) on the release module RM02 again – the image is
saved
5.4 White balance
The Haag‑Streit IM 600 is optimized for maximum image quality with the Haag‑St‑
reit slit lamp BQ 900 and BP 900. The image quality is dependent, among other
things, on the correct calibration of the color tones to the respective slit lamp illumi‑
nation. We recommend performing a white balance in order to improve the image
quality and achieve a realistic color reproduction.
5.4.1 Slit lamp preparation
1. Turn on the slit lamp
2.
Filter position ‹open› (no filter)
3.
Set magnification to 16 ×
4. Completely open the slit diaphragm
5. Connect the diffuser upstream
6. Position the Haag‑Streit greycard in front of the slit lamp and use it for focusing
7. The brightness of the slit lamp's illumination should be set in such a way that
the greycard's structure is clearly discernible.
5.4.2 Conducting a white balance
8. Start the ‹EyeSuite Imaging› software
9. Activate the intensity auto mode
10. Open the ‹White balance› application
11. Start the ‹White balance› by activating the ‹Calibration› function
WARNING!
To achieve an optimal result during the white balance, the image must
be homogeneously illuminated.
Set white balance greycards
REF 1021485