F.  Information and manufacturer's declaration  

 

concerning electromagnetic compatibility  

 (EMC)

F.1 General

The Imaging Module IM 600

 system fulfills the requirements on electromagnetic 

compatibility according to EN 60601‑1‑2:2007 (IEC 4. Edition) + EN 60601‑1‑2:2015 

(IEC 4. Edition). The instrument is built so that the generation and emission of elec‑

tromagnetic interference is limited to the extent that other devices are not disturbed 

in their use in accordance with the regulations and so that the instrument itself is 

suitably immune to electromagnetic interference.

WARNING!

Avoid damages due to high electrostatic discharges (ESD). Electrosta‑

tic discharges with voltages exceeding 6 kV to some parts of the slit 

lamp like joystick or metallic parts on the instrument base may influen

ce the instrument.

• 

The communication between Imaging Module and PC could be inter‑

rupted, which would require a restart of the EyeSuite software.

WARNING!

• 

Electrical medical devices and systems are subject to special EMC 

measures and must be installed in accordance with the EMC instruc-

tions contained in this accompanying document.

The operation of other lines or equipment than those listed may lead 

to higher emissions or may reduce the device's resistance to interfe-

rence.

Third-party devices may only be connected in compliance with the 

EN 60601-1 standard.

 

 

F.2  Emitted interference (standard table 1)

The information is based on the requirements of EN 60601‑1‑2:2007 (IEC 3

rd

 edition) and EN 60601‑1‑2:2015 (IEC 4

th

 edition).

Guidance and manufacturer's declaration – electromagnetic emissions

This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi-

ronment.

Emission test

Compliance

Electromagnetic environment - guidance

RF emissions CISPR 11

Group 1

This product uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class A

This product is suitable for use in all establishments, including domestic establishments and those directly connected to 

the public low-voltage power supply network that supplies buildings used for domestic purposes.

Emission of harmonics 

according to EN 61000-3-2

Class A

01-IFU_IM600-7220506-04050_eng.indd   15

11.06.2019   16:03:07

Summary of Contents for IM 600

Page 1: ...A HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 DOC No 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Imaging Module IM 600 Accessory for slit lamps BQ 900 and BP 900 5 edi...

Page 2: ...urpose of use This device is an accessory for Haag Streit slit lamps BQ 900 or BP 900 which can be used to produce digital photographs and videos for documentation of the eye Contraindication There ar...

Page 3: ...igger 10 5 4 White balance 10 5 4 1 Slit lamp preparation 10 5 4 2 Conducting a white balance 10 5 5 Software Help menu error messages 11 5 6 LED display illumination head 11 5 7 LED display power sup...

Page 4: ...hPa 95 Use Temperature Air pressure Relative humidity from from from 10 C 800 hPa 30 to to to 35 C 1060 hPa 90 Air pressure Relative humidity from from 500 hPa 10 to to 1060 hPa 95 Storage Temperature...

Page 5: ...xpressly disclaims liability for incidental or consequential damage resulting from the use of the product This product is covered by a limited warranty granted by your seller This product is covered b...

Page 6: ...umination see separate instructions for use 3 Fiber optic line for periphery or background illumination 4 Headrest see separate instructions for use 5 Rail cover 2 4 Camera module CM02 CM02BP 6 Camera...

Page 7: ...at the corner opposite the black surface Position the sticker against the right roller rail 18 and the gliding plate 19 Press firmly on the white black surface press away any air bubbles Carefully te...

Page 8: ...connection plug If the middle LED lights up red during operation the two pole connection plug 39 is connected incorrectly Disconnect the device from the pow er grid Remove the cover on the upper part...

Page 9: ...se module RM02 51 RM02 operational control LED 52 On Off key 53 Rotating knob periphery or back ground illumination 54 Rotating knob slit illumination 55 Selector key A 56 Selector key B 57 Release ke...

Page 10: ...ease key 57 on the release module RM02 again the image is saved 5 4 White balance The Haag Streit IM 600 is optimized for maximum image quality with the Haag St reit slit lamp BQ 900 and BP 900 The im...

Page 11: ...illumination Polarity Slit lamp illumination a b c Operating status Standby mode Green short flashes Green short flashes Normal operation Slit and periphery or background illumi nation on Green Green...

Page 12: ...ached Red flashing Red flashing E4 No communication between power supply and illumination Contact your Haag Streit representative Red flashing 2 Red flashing 2 E6 General error Send PS LED to the appr...

Page 13: ...M if third party applications such as patient ad ministration software are to be used alongside EyeSuite Use 2 RAM module Hard disk At least 500 GB NTFS data system Graphics Graphics card with at leas...

Page 14: ...nst liquids A Appendix A 1 Accessories consumables spare parts upgrade Components REF Set white balance greycards 1021485 Power supply ICCNEXERGY 1022106 Computer cable USB 3 0 1022373 Release module...

Page 15: ...be installed in accordance with the EMC instruc tions contained in this accompanying document The operation of other lines or equipment than those listed may lead to higher emissions or may reduce th...

Page 16: ...t of a typical commercial or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 1 kV for symmetrical voltages Mains power quality should be that of a typical commercial or hospital...

Page 17: ...aration distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference...

Page 18: ...output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 7 80 MHz 800 MHz D 1 2 800 MHz 2 5 GHz D 2 3 0 01 0 1 1 10 100 For transmitters with a nominal outpu...

Page 19: ...EUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA HAAG STREIT AG 3098 Koeniz Switzerland HS Doc No 1500 7220506 04050 2019 06 01 IFU_IM600 7220506 04050_eng indd 19 11 06 2019 16 0...

Page 20: ...u have any further questions please contact your Haag Streit representative at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerla...

Reviews: