Check position 1.95
Displace the calibration marking 2 on the measuring drum down-
wards against the index marker by the width of the marking.
When the feeler arm is slightly moved in the direction of the
free movement zone, it should move independently against
the stop piece in the direction of the examiner.
Check position 2.05
Displace the calibration marking 2 on the measuring drum up-
wards against the index marker by the width of the marking.
The feeler arm should move against the stop piece on the
side of the patient.
Check at measuring drum setting 6
Turn the weight bar to scale calibration 6, the longer part shows
in the direction of the patient.
Check position 5.9
The check point is 5.9. The calibration marking 6 on the measu-
ring drum is to be turned 1/2 an interval downwards. The feeler
arm should move towards the examiner.
Check position 6.1
The check point is 6.1. The calibration marking 6 on the measu-
ring drum is to be turned 1/2 an interval upwards. The feeler
arm should move in the direction of the patient.