2. Introduction
The applanation tonometer functions according to the 'Goldmann method': mea‑
surement of the pressure required to maintain a uniform applanation of the surface
of the cornea. The precise measurement of the applanation surface is conducted
with the slit lamp at 10x magnification.
The pressure measurement is conducted on a slit lamp with the patient sitting down
and is a routine test carried out as part of the standard slit lamp microscopy.
2.1 Overview
1. Measuring prisms (application part)
2. Sensor arm
3. Battery compartment lid
4. Battery compartment locking mechanism
5. Type plate (bottom side)
6. Function key
7. Insert for control weight
8. Digital display
9. Knob
10. Bluetooth adapter (optional)
1
2
3
4
5
6
7
8
9
2.2 System components
The Applanation Tonometer system is composed of the following devices:
•
Applanation Tonometer
•
Bluetooth adapter (optional)
•
PC or laptop
2.3 Bluetooth adapter (optional)
There is an optional Bluetooth adapter available to transfer the measured data directly
to the patient management of the Eyesuite software. The activation of the Bluetooth
function is described in Chapter 4.6. The description of the software installation can be
found in the EyeSuite Setup Manual, the usage instructions for the Bluetooth adapter
and the description of the error messages are included in the Help section of the soft‑
ware. The help can be opened via the F1 key or in the [?] ‑ [Help] menu.
3. Appliance assembly / installation
WARNING!
Installation, repairs and modifications may only be performed by
trained specialists.
3.1 Model R
A pivot is required for use of the Applanation Tonometer model R o n the BM 900 slit
lamp. It is assembled as follows:
1. Unscrew the mounting screw found at the top, on the microscope's middle cyl‑
inder housing.
2. Leave the spring in the aperture.
3. Position and screw on tightly the pivot for the tonometer.
4. Place the tonometer bearer arm on the pivot's pin and swivel the tonometer to
the right until it snaps into position. The tonometer will remain in this position
even when not in use.
5. When using the BI 900, BM 900 and BM 900 V, the stop peg must be assem‑
bled (a), and when using the BC 900 or BD 900, it must be removed (b).
(b)
(a)
3.2 Model T
The Applanation Tonometer is tucked into one of the holes on the horizontal guide
plate over the slit lamp axis using the peg from the tonometer base.
3.3 Model BQ
Connect the tonometer at the provided interface to the right side of the microscope
arm.
NOTE!
The cover screw on the left side of the arm is loosened by turning it in
a clockwise direction using the screwdriver (applies only for late tonom‑
eter deliveries).