Test at 60 mm Hg
Set the weight rod to 60 on the scale so that the longer part points to the examiner.
Check position 59
The sensor arm must not move from the free‑movement area toward the limit stop
in the direction of the examiner before the display shows 59.0 mm Hg or after the
display shows 61.0 mm Hg.
Check position 61
The sensor arm must not move from the free‑movement area toward the limit stop
in the direction of the patient before the display shows 61.0 mm Hg or after the dis‑
play shows 59.0 mm Hg.
8.4 Changing the battery
1. Push the HAAG‑STREIT logo (HS) on the side of the digital display carefully
downward and remove the battery lid by tilting it backward.
2. Pull the tab at the bottom of the battery compartment carefully. The batteries
are released from the holder and can be easily removed.
WARNING!
•
Pay attention to the polarity when inserting the batteries.
•
Always replace all the batteries with new ones
•
Only use type LR03 (AAA) batteries.
•
Never mix old and new batteries.
2.
1.
WARNING!
Pay attention to the polarity when inserting the batteries.
3. Pay attention to the correct order as written in the compartment when inserting
the batteries.
4. Once you have replaced the batteries, close the lid again carefully. Insert the
centering tab at the bottom of the battery lid into the corresponding groove in
the housing and close the lid by tilting it upward.
5. Gently press the lid upward against the housing and lock the battery compart‑
ment by pushing the HAAG‑STREIT logo back up again.
NOTE!
Dispose of used batteries properly.