5.8.3 Position too high/low
Only part of a semicircle in the upper half (10)
Measuring prism not centered on the eye; eye far too high.
10
Using the control lever, move the slit lamp upwards.
Nearly entire circle above, incomplete circle below (11)
Measuring prism not centered on the eye; eye still too high.
11
Using the control lever, move the slit lamp upwards.
Two incomplete circles, the larger one on top (12)
Measuring prism not centered on the eye; eye still too high.
12
Using the control lever, move the slit lamp upwards.
Correct setting! (13)
Two semicircles appear exactly in the middle of the eyepiece.
13
5.8.4 Incorrect pressure
The outside borders of the fluorescein bands are in contact
with each other (14)
Not enough pressure.
14
Increase the pressure slightly by rotating the knob on the
tonometer.
Fluorescein bands are superimposed to form a band (15)
Pressure slightly too low.
15
Increase the pressure slightly more by rotating the knob
on the tonometer.
Bands are no longer in contact (16)
Pressure is too high.
16
Lower the pressure by rotating the tonometer knob in the
opposite direction.
Correct setting! (17)
The inside borders of the fluorescein bands are contacting each
other.
17
6. Software / Help menu / Error messages
The software's help section contains instructions and help for performing an exami‑
nation and descriptions of the error messages. The help can be opened via the F1
key or in the [?] ‑ [Help] menu.
WARNING!
The software must be installed by trained personnel in accordance with
separate installation instructions.
6.1 Error messages on the tonometer display
Description
Measures
The measurement is outside the
permitted tolerances.
The position of the sensor arm is
defective.
If one of these errors appears per‑
manently, send the device to the
corresponding service branch.
The minimum operating voltage
has not been achieved; the device
switches off.
Replace the dead batteries.
Battery voltage too high.
Incorrect battery type fitted.
Insert the correct batteries.