GDM-834X Series User Manual
150
DC Voltage
Range
Resolution
Full Scale
Accuracy
(1 year 23
˚
C ±5
˚
C)
Input
Resistance
500mV
10
μ
V
510.00
0.02%+4
10M
Ω
or
>10G
Ω
5V
100
μ
V
5.1000
10M
Ω
or
>10G
Ω
50V
1mV
51.000
11.1M
Ω
500V
10mV
510.00
10.1M
Ω
1000V
100mV
1020.0
10M
Ω
* When the input value exceeds the full scale of the selected range, the display will
show -OL- (over load) on the display.
* The specifications are guaranteed to an input voltage of 1000V. A beeping alarm
will go off when the input voltage is higher than 1000V.
* Input protection of 1000V peak on all ranges.
DC Current
Range
Resolution Full Scale
Accuracy
(1 year 23
˚
C ±5
˚
C)
Shunt
Resistance
Burden
Voltage
500
μ
A
10nA
510.00
0.05%+5
100
Ω
0.06V max
5mA
100nA
5.1000
0.05%+4
100
Ω
0.6V max
50mA
1
μ
A
51.000
0.05%+4
1
Ω
0.14V max
500mA
10
μ
A
510.00
0.10%+4
1
Ω
1.4V max
5 A
100
μ
A
5.1000
0.25%+5
10m
Ω
0.5V max
10 A
1mA
12.000
0.25%+5
10m
Ω
0.8V max
* 500
μ
A~500mA range has a 3.6V voltage limit protection and 0.5A fuse
protection. And 10A range has a 12A fuse protection.
* When the input value exceeds the full scale of the selected range, the display will
show -OL- (over load) on the display.
* The specifications are guaranteed to an input of 10A. A beeping alarm will go off
when the input value is higher than 10A.