COMMAND OVERVIEW
117
CONFigure:CURRent:AC
Sets measurement to AC Current on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:AC 50e-2
Sets the measurement mode to ACI with a 500mA range.
CONFigure:CURRent:DCAC
Sets measurement to DC+AC Current on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:DCAC 50e-2
Sets the measurement mode to DC+AC Current with a 500mA
range.
CONFigure:RESistance
Sets measurement to 2W Resistance on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:RES 50e3
Sets the range to 50k
Ω
.
CONFigure:FREQuency
Sets measurement to Frequency on the first display and specifies
the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:FREQ MAX
Sets the frequency measurement range to max.