COMMAND OVERVIEW
119
CONFigure:RANGe?
Returns the current range on the first display.
Return Parameter:
DCV: 0 .5(500mV), 5(5V), 50(50V), 500(500V), 1000(1000V)
ACV: 0.5(500mV), 5(5V), 50(50V), 500(500V), 750(750V)
ACI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A),
10(10A)
DCI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A),
10(10A)
RES: 50E+1(500
Ω
) 50E+2(5k
Ω
), 50E+3(50k
Ω
), 50E+4 (500k
Ω
),
50E+5(5M
Ω
), 50E+6(50M
Ω
)
CAP: 5E-9(5nF), 5E-8(50nF), 5E-7(500nF), 5E-6(5μF), 5E-5(50μF)
CONFigure:AUTO
Sets Auto-Range on or off on the first display.
Parameter: ON | OFF
Example: CONF:AUTO ON
CONFigure:AUTO?
Returns the Auto-Range status of the function on the 1
st
display.
Return Parameter: 0|1, 1=Auto range, 0=Manual range
Secondary Display: CONFigure2 Commands
CONFigure2:VOLTage:DC
Sets measurement to DC Voltage on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:VOLT:DC 5
Sets the voltage range to 5 volts.