APPENDIX
153
Capacitance
Range
Resolution
Full Scale Test Current
Accuracy
(1 year 23
˚
C ±5
˚
C) [1]
5nF: 0.5~1nF [2]
0.001nF
5.100
8.3
μ
A
2.0%+20
5nF: 1~5nF [2]
2.0%+10
50nF: 5~10nF [2]
0.01nF
51.00
8.3
μ
A
2.0%+30
50nF: 10~50nF [2]
2.0%+10
500nF
0.1nF
510.0
83
μ
A
2.0%+4
5
μ
F
1nF
5.100
0.56mA
50
μ
F
10nF
51.00
0.83mA
[1] For the 5nF ~ 50
μ
F
range, make sure that the input is greater than 10% of the
range.
[2] Need to use the REL function.
* Input protection of 500V peak on all ranges.
Frequency
Measurement Range
Accuracy (1 year 23
˚
C ±5
˚
C)
10Hz ~ 500Hz
0.01%+5
500Hz ~ 500kHz
0.01%+3
* AC + DC measurements do not allow frequency measurements.
* Input protection of 1000V peak on all ranges.
Voltage Measurement Sensitivity
Range
Minimum Sensitivity (RMS sine wave)
10~100kHz
100K~500kHz
500mV
35mV
200mV
5V
0.25V
0.5V
50V
2.5V
5V
500V
25V
uncal
750V
50V
uncal
Current Measurement Sensitivity
Range
Minimum Sensitivity (RMS sine wave)
30~20kHz
500
μ
A
35
μ
A
5mA
0.25mA
50mA
2.5mA
500mA
25mA
5 A
0.25A(<2kHz)
10 A
2.5A(<2kHz)
www.
GlobalTestSupply
.com
Find Quality Products Online at: