OPERATION
41
Select the Current Range
The current range can be set automatically or manually.
Auto Range
To turn the automatic range selection On/Off,
press the AUTO key.
Manual Range
Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.
Selectable
Current Ranges
Range
Resolution Full scale
500µA
10nA
510.00μA
5mA
100nA
5.1000mA
50mA
1μA
51.000mA
500mA
10μA
510.00mA
5A
100μA
5.1000A
10A
1mA
12.000A
Note
For further details, please see the specifications on
page 149.
Note
DC currents with AC components cannot be
accurately measured if the DC+AC component
exceed the ADC dynamic range for the selected DC
range. Any current exceeding the ADC dynamic
range will be clipped at the upper/lower range
limit. Under these conditions the range that is
chosen with the Auto range function may be too
small.