PPX Series User Manual
80
OVP Level
It sets over voltage protection setting for the
select step.
OVP Level
5% ~ 110% rated voltage
OCP Level
It sets over current protection setting for the
select step.
OCP Level
5% ~ 110% rated current
Bleeder
It enables or disables discharge loop control for
the select step.
Bleeder
None, ON, OFF
V/I Slew Rate
It sets High Speed Priority and Slew Rate
Priority of CV and CC modes for the select step.
V/I Slew Rate
CVHS
It utilizes the fastest slew rate of CV
mode. Refer to page 51 for more
details.
CCHS
It utilizes the fastest slew rate of CC
mode. Refer to page 55 for more
details.
CVLS
It utilizes the user-configured slew
rate of CV mode. When this option is
selected, go to configure the R_V slew
Rate (rising) and F_V slew rate
(falling) settings, respectively. Refer
to page 51 for more details.
CCLS
It utilizes the user-configured slew
rate of CC mode. When this option is
selected, go to configure the R_C slew
Rate (rising) and F_C slew rate
(falling) settings, respectively. Refer
to page 55 for more details.