OPERATION
49
Setting Range
Model
OCP
OVP
UVL
PPX-1005
0.25~5.5
0.5~11
0~10.5
PPX-2002
0.1~2.2
1~22
0~21.0
PPX-2005
0.25~5.5
1~22
0~21.0
PPX-3601
0.05~1.1
1.8~39.6
0~37.8
PPX-3603
0.15~3.3
1.8~39.6
0~37.8
PPX-10H01
0.05~1.1
5~110
0~105
Note
The UVL setting range is from 0% ~ 105% of the rated
output voltage. It depends on Voltage Limit On/Off to
activate/deactivate UVL setting.
The OVP setting range is from 5% ~ 110% of the rated
output voltage.
The OCP setting range is from 5% ~ 110% of the rated
output current.
Setting the
Delay Time
4.
Scroll the knob key to move
between OCP Delay fields. Click
the knob key to enter each field,
respectively. Scroll the knob key to
adjust value, along with the arrow
keys to change among digits
followed by clicking the knob key
to confirm set value.
Setting Range
OCP Delay
0.05~2.500 s