1.09
Power supply
GH1 is equipped with a battery that requires regular recharging. The power
supply for the transformer and the charging station must be connected by
Guldmann Service Team or by a qualified engineer.
The transformer supplied must
always
be used.
Safety concerning static electricity (ESD)
Service technicians and installers
must
use an ESD-safety package
consisting of a mat, a ground wire, and a bracelet.
The technician/installer connects the mat to a grounding point, for instance a
radiator or a water pipe. The technician/installer must then put on the bracelet
and connect it to the mat. If it is not possible to find a grounding point, the
mat and the bracelet must be used as a minimum.
Only then is it allowed to work with the PC Board or components where it
is possible to come into contact with the PC Board.
Class II equipment
Mobile equipment is class II epuipment (marked with double-encassed
symbol) and can be connected to the mains direct by the user.
Equipment is disconnected from Supply Mains by detaching the mains plug
from the wall outlet.
8
© Guldmann GB/US-1706/12/2016 • # 550865_14
© Guldmann GB/US-1706/12/2016 • # 550865_14