18
© Guldmann GB/US-1706/12/2016 • # 550865_14
© Guldmann GB/US-1706/12/2016 • # 550865_14
2.04
Safety functions
The emergency stop and emergency lowering device should only be
used in an emergency. In the event that it is necessary to use the safety func-
tions, the fault must be identified and rectified before the GH1 is taken into
use again. Please contact your supplier.
Emergency stop and lowering strap
The red strap has the following functions:
• One pull: Emergency stop is activated.
• Constant pull: Emergency lowering is activated.
Emergency stop
If the GH1 does not stop/react to the hand control when the GH1 is in
use, pull the red strap and the lifting/lowering functions (except emergency
lowering) are deactivated.
When the emergency stop is activated, the hoist will not function. The green
lamp is switched off.
Reset emergency stop
Reset the emergency stop by pressing the yellow
button on the bottom of the hoist.
The yellow button that appears when the emer-
gency stop is pulled, must be pressed manually
before the GH1 is ready for use.
After deactivating the emergency stop, activate the
hand control twice.
Emergency lowering function, electric
If the GH1 fails, the electrical emergency lowering function is used to lower
the user safely. The emergency lowering function is operated by a constant
pull on the red strap that is used for the emergency stop.
When releasing the red strap, the emergency lowering function will be
replaced by the emergency stop.