CyScan Fuse Information
The CyScan system contains two replaceable fuses, one for the Live and the other for the
Neutral. These are located on the connector board as shown below.
Connector Board Fuse
Please refer to the Installer’s Guide for details on how
to access the connector board.
To remove a fuse from its housing, fit a flat-head screwdriver into
the slot in the insert - then push it in and turn it anti-clockwise until
the insert releases and can be pulled out.
Reverse the above procedure to replace the fuse.
Both fuses have the same specification.
Live
Neutral
Ensure the replacement fuse has the same specification as the
original fuse.
CAUTION: DOUBLE POLE/NEUTRAL FUSING
Voltage Supply
230v
110v
Normal Operation
630mA
630mA
Low temperature operation
1A
2A
The fuses should be 20mm anti-surge glass fuses rated for 250V.
The current rating should be determined by the table below.
Normal operation is where the sensor is operated at -10°C or
above.
Operating at below this temperature we have additional internal
heaters which warm the unit at start-up