background image

English (GB)

19

6.3.2 Setting the water flow rate

The water flow rate is shown at flowmeter (6), and can be 
adjusted with water flow adjustment valve (8).

* Observe target value of specific gravity (SG). See section 

3.4 Typical unit set points

.

6.4 Start-up and shutdown procedures

6.4.1 Start-up

1. Open water isolating valve (16a) and set pressure reducing 

valve (10) to show 3-4 bar on gauge.

2. Make sure that the brine tank (A) is prepared according to 

sectio

6.1 Brine tank

.

3. Open isolating valves (16b), (16c) and (16d).
4. At initial start-up or if the system was shut down for more than 

4 weeks, regenerate the water softener manually. See section 

6.2.1 Manual regeneration

.

5. Switch on power supply with mains switch (29). A start-up 

timer will be initiated at display (28).

6. Adjust the water flow adjustment valve (8) to set the water 

flow rate given in section 

3.4 Typical unit set points

 at the 

flowmeter (6).

7. Set the brine pump (3) to achieve a brine solution with a 

specific gravity (SG) of 1.02 using the Selcoperm test kit.

– See section 

6.5.2 Checking the salinity of the electrolyte

.

– The brine pump (3) is regulated manually.

6.4.2 Shutdown

Short term shutdown (up to 4 weeks)

Use the [LEFT] key on control (28) to shut down the Selcoperm 
system for a short term. See section 

5. Operation

.

Shutdown up to six months

1. Switch off power supply at switch (29).
2. Isolate system from water supply with isolating valve (16a).

Shutdown for periods over six months

1. Switch off power supply at switch (29).
2. Fit hose from soft water sample valve (F) to electrolyte sample 

valve (G).

3. Fit hose to product sample valve (H) and direct it to a suitable 

discharge point.

4. Open product sample valve (H), electrolyte sample valve (G) 

and soft water sample valve (F).

5. Flush system thoroughly for a period of 5-10 minutes.
6. Close soft water sample valve (F). The degassing column (2) 

drains automatically through open product sample valve (H).

7. Close electrolyte sample valve (G).
8. Remove hose from soft water sample valve (F) and direct 

hose from electrolyte sample valve (G) to a suitable discharge 
point.

9. Open electrolyte sample valve (G) to drain electrolyser 

cell (1).

10. Close all sample valves and remove hoses.
11. Isolate system from water supply with isolating valve (16a).

For further advice please contact Grundfos.

6.5 Checks during operation

6.5.1 Checking the water hardness

A properly functioning water softener (4) is crucial for the 
efficiency and longevity of the Selcoperm system. High levels of 
calcium carbonate in the electrolyte will lead to a visible 
precipitation on the electrodes of electrolyser cell (1).
We recommend to perform this check with the Grundfos 
Selcoperm test kit. See section 

7.3.1 Test kit

.

1. Take a sample of water from soft water sample valve (F).
2. Perform the test according to the test kit operating 

instructions.

6.5.2 Checking the salinity of the electrolyte

Checking the specific gravity (SG) of the electrolyte

We recommend to perform this check with the Grundfos 
Selcoperm test kit. See section 

7.3.1 Test kit

.

1. Flush a clean and dry measuring cylinder with electrolyte from 

the electrolyte sample valve.

2. Fill the measuring cylinder with a sample from the electrolyte 

sample valve.

3. Place the hydrometer in the measuring cylinder.
4. Take the reading from the hydrometer.

– Required SG reading see section 

3.4 Typical unit set points

.

Adjusting the SG

The SG of the electrolyte is controlled by the amount of brine 
injected into the softened water. The brine flow is controlled by a 
digital dosing pump (3). For the operation of the pump, see the 
manual of the pump. Increasing the flow of brine injected into the 
softened water will increase the SG, and restricting the flow will 
lower the SG.

Type

Factory-set 

brine flow [l/h]*

Factory-set 

water meter flow 

[l/h]

Total product 

flow [l/h]

125

1.7 - 2.1

19

20.7 - 21.1

250

3.4 - 4.2

38

41.4 - 42.2

500

6.8 - 8.2

75

81.8 - 83.2

1000

12 - 15.5

140

152 - 155.5

2000

24-30

285

309-315

Note

The values given above have to be kept and may 
require adjustment at site due to salt quality, water 
pressure, temperature and mains water flow rate.

Note

Product dosing pumps should be set to give the 
desired chlorine at the point of injection.

Note

The frequency of the tests depends on the specific 
application, and must be determined individually by 
the customer.

Note

The salinity can be checked either by measuring the 
conductivity with the Grundfos conductivity 
measuring device or by measuring the specific 
gravity with the Grundfos Selcoperm test kit.
See section 

7.3.1 Test kit

.

For details on the conductivity measuring device, see 
data booklet: http://net.grundfos.com/qr/i/98721404

Summary of Contents for Selcoperm SES 125-2000

Page 1: ...elcoperm SES 125 2000 Electrochlorinator Valid for SES 125 250 500 1000 built from week 37 2015 Valid for SES 2000 built from week 20 2014 Installation and operating instructions GRUNDFOS INSTRUCTIONS...

Page 2: ...em via a display with control and display elements 1 General safety instructions 2 1 1 Purpose of this document 2 1 2 Symbols used in this document 2 1 3 Users target groups 2 1 4 Obligations of the o...

Page 3: ...guaranteed for a period of two calendar years from installation and commissioning provided optimum operating conditions are as detailed in this manual The maximum period for full guarantee is 36 month...

Page 4: ...al It is the duty of the operator to ensure that he has obtained information on all safety precautions relevant to the type of task he is undertaking and to ensure that they are adhered to Grundfos re...

Page 5: ...ood switch F Soft water sample valve 7 Non return valve 19 Air dilution fan G Electrolyte sample valve 8 Water flow adjustment valve 20 Flowmeter switch H Product sample valve 9 Degassing column level...

Page 6: ...ment valve 21 Floater valve G Electrolyte sample valve 9 Degassing column level sensor 22 Control panel H Product sample valve 10 Pressure reducing valve 23 Control lamp of mains supply I Product over...

Page 7: ...ank M When the product reaches the stop fill level 25b in the product tank M the process is stopped The process is restarted when the product level in the tank M falls below the start fill level The p...

Page 8: ...minal g h 125 110 250 220 500 450 1000 900 2000 1800 Connection I imperial M metric Supply voltage frequency H 110 120 V 50 60 Hz G 220 240 V 50 60 Hz K 380 415 V 50 60 Hz Software language GB English...

Page 9: ...2 Dimensional drawing Fig 5 Dimensional drawing Type 125 250 500 1000 2000 Gross weight kg 299 309 319 334 349 Net weight kg 160 170 180 195 210 TM04 1594 3011 712 457 104 300 1570 1112 215 170 245 3...

Page 10: ...to 40 Permissible mains water operating temperature C 10 to 20 Transport and storage temperature drained and not connected C 5 to 50 Max altitude above sea level m 2000 System frame PP Chamber covers...

Page 11: ...k and collecting tray X30 7 X30 7 GND BN 24 VDC WH Signal GN Tray flood switch of product tank Product tank level sensor Single phase Three phase Mains power supply To customer telemetry Mains power s...

Page 12: ...ray No liquid in collecting tray Remote stop signal System stopped System running Emergency stop signal System stopped System running Data 125 125 250 500 1000 2000 single phase single phase three pha...

Page 13: ...ng Horizontal ventilation pipework has to be installed with a constant upwards incline towards a safe discharge point min 2000 min 700 min 150 1 Duty standby dosing pumps 5 Pressure level sensor 2 Pro...

Page 14: ...017 3 3 3 1 1 2 1 Soft water inlet 20 mm 2 Minimum gravel level 3 Brine outlets 20 mm Note Requirements and capacities of brine tank see section 3 6 Brine tank capacities TM04 1504 3011 6 5 4 3 2 2 1...

Page 15: ...VC pipework to the PE pipework Make sure that the product line from the electrolyser to the product tank is installed horizontally or downwards See fig 8 The table below shows the connection sizes for...

Page 16: ...lues Press ESC to exit the edit mode Once all values are set press ESC to exit service menu 1 and return to main display screen in main menu Warning Before making any electrical connections switch off...

Page 17: ...SC ESC ESC GRUNDFOS INFORMATION WWW GRUNDFOS COM ENTER PASSWORD 01987 1 AIR FLOW X m hr GENERATING TIME X hrs SERVICE MENU 1 X AMPS X VOLTS VOLTAGE ALARM ALARM TIMER X sec AIR FLOW m hr TANK HEIGHT X...

Page 18: ...d turn the screw clockwise until pin 2 is positioned in regeneration start range 4 for regenerating the right side of the softener During regeneration pin 2 moves clockwise Wait until pin 2 has left t...

Page 19: ...th isolating valve 16a For further advice please contact Grundfos 6 5 Checks during operation 6 5 1 Checking the water hardness A properly functioning water softener 4 is crucial for the efficiency an...

Page 20: ...ing time The alarm timer must be reset after completion of testing After the triggering of each alarm the electrochlorinator will need to be reset at control 28 6 8 1 Internal tray flood alarm The int...

Page 21: ...Safety installations which have been disabled during maintenance must be enabled again immediately after maintenance Warning Danger of personal injury from chlorine gas Sodium hypochlorite solution te...

Page 22: ...leaning If necessary clean all system surfaces with a moist and clean cloth 7 8 Acid cleaning 1 Switch off power supply at switch 29 and isolate system from the mains voltage 2 Connect hose to product...

Page 23: ...erature of electrolyser cell 1 to high CELL TEMP Mains water temperature to high Check and rectify High ambient temperature Ambient temperature must not exceed 40 C Faulty temperature sensor Replace t...

Page 24: ...or 98056760 Electrolyser cell SES 2000 14 98058479 Brine valve water softener 3 97721401 Brine pump SES 125 250 500 15 98058478 Brine injection valve 97722252 Brine pump SES 1000 2000 17 98056894 Inte...

Page 25: ...ller language Russian 2 95727184 Temperature sensor 3 98710813 Electrolyser power supply SES 125 98710815 Electrolyser power supply SES 250 98710816 Electrolyser power supply SES 500 98710817 Electrol...

Page 26: ...until spring 2009 TM04 1506 3311 Pos Product No Description 1 98056828 Side entry valve 2 95727187 3 4 lateral filter TM04 1595 3311 Pos Product No Description 1 98056894 Tray flood switch 2 95721859...

Page 27: ...Grundfos company or service workshop 11 Photos Fig 20 Correct installation of Venturi tee Fig 21 Typical installation Warning Before dismantling the system must be completely flushed with water in or...

Page 28: ...English GB 28 Fig 22 Installation pipework Fig 23 Safe ventilation point Subject to alterations TM04 1514 1210 TM04 1515 1210...

Page 29: ...niejsza dotyczy jest zgodny z nast puj cymi dyrektywami Rady w sprawie zbli enia przepis w prawnych pa stw cz onkowskich PT Declara o de conformidade UE A Grundfos declara sob sua nica responsabilidad...

Page 30: ...Declaration of conformity EAC 30 Declaration of conformity EAC 2 Selcoperm SES SEP 004 2011 010 2011 020 2011 TC RU C DK 30 01060 05 11 2019 153032 1 1 2016 143581 188...

Page 31: ...an GRUNDFOS Pumps K K 1 2 3 Shin Miyakoda Kita ku Hamamatsu 431 2103 Japan Phone 81 53 428 4760 Telefax 81 53 428 5005 Korea GRUNDFOS Pumps Korea Ltd 6th Floor Aju Building 679 5 Yeoksam dong Kangnam...

Page 32: ...3971 The name Grundfos the Grundfos logo and be think innovate are registered trademarks owned by Grundfos Holding A S or Grundfos A S Denmark All rights reserved worldwide Copyright Grundfos Holding...

Reviews: