English (GB)
45
"OFF CAL" appears in the display.
1. Press the "Up" button.
– "ON CAL" appears in the display.
2. Press the "Start/Stop" button.
– Calibration is started. The LED flashes green
and "Run" flashes in the display.
•
200 strokes are performed by default.
The number of strokes performed appears in the
display.
•
Press "Start/Stop" to stop at any stroke value.
– The LED lights up red.
3. Press the "Start/Stop" button.
– The current calibration value is displayed
(not with replacement circuit boards!).
4. Use the "Up" and "Down" buttons to enter the
calibration value (gauged volume in ml).
•
Press the "Menu/Info" button (confirm the setting
and move on to the next menu item), or
•
press the "Start/Stop" button (confirm the setting
and close the second function level).
Fig. 44
Calibration
Note
The calibration process does not
change the setting for the
assignment/weighting of the current
input and output for the flow rate.
If a current input or output is used,
check after calibration to determine
whether a new current weighting is
required.
TM
03 66
18 450
6
110.0
Stop
99
Run
112.0
Stop
Down
Up
Stop
Start
87
Stop
200
Stop
Stop
Start
CAL
CAL
CAL
CAL
CAL
ml
Stop
Start
OFF
Up
Down
OFF
CAL
CAL
On
CAL
Menu
Info
Menu
Info
Menu
Info
Menu
Info
Flashing
Green LED flashing
Process 200 strokes
Red LED
Red LED
Manual stop at user-defined stroke number
Value of manufacturing calibration - calculated for stroke
number of actual calibration
Input of actual ascertained calibration
volume