2.3
Start your Measurement by clicking
Start logging
A Blue Light will flash from the Base Station logger to indicate that it is in logging mode.
NOTE:
If you wish to check that measurements are being taken, then click
Real-time
measurement
on the Home Page to see the incoming signals. Readings may not show up
straight away if node is set to send measurements ever 10 mins however.
You may now disconnect the Base Station from your PC. It is recommended that you keep
your Base Station charging during measurements, as the Autonomous Battery Life is
approximately 48 hours; Below the minimum required 72 hours for an ISO 9869 conforming
measurement.
BASE STATION PLACEMENT RECOMMENDATIONS
The ideal placement for the Base Station is in a direct line-of-sight with the measurement
nodes.
greenTEG observed an
indoo
r range of
50m
with the signal travelling through concrete
walls, and an
outdoor
signal range of over
200m
.
This Range can worsen with the amount of obstacles between the measurement
nodes and the Base Station. Range is particularly worsened by
Thick Walls
, and those
made from
Concrete
.
If Measurements are being taken on Different Floors/Stories, avoid placing the Base
Station directly above/below Nodes.
STARTING YOUR MEASUREMENTS
11
Summary of Contents for GO MEASUREMENT SYSTEM II
Page 1: ...gOMS II greenTEG GO MEASUREMENT SYSTEM II USER MANUAL...
Page 2: ...CONTENTS I 01 02 03 04 OVERVIEW SET UP ANALYSING MEASUREMENTS APPENDIX...
Page 9: ...01 02 03 04 OVERVIEW SET UP ANALYSING MEASUREMENTS APPENDIX CONTENTS 7...
Page 13: ...CONTENTS 12 01 02 03 04 OVERVIEW SET UP ANALYSING MEASUREMENTS APPENDIX...
Page 18: ...CONTENTS 17 01 02 03 04 OVERVIEW SET UP ANALYSING MEASUREMENTS APPENDIX...
Page 22: ...Appendix C Detailed explanations of the parameters used in the report 4 3 21...