5) Pulse Rate Control
Pulse rate is adjustable. Its value is visible on
the LCD. This buttons determine how many
electrical impulses are applied through the skin
every second. By pressing these controls, the
number of current impulses per second (Hz)
for both channels can be continually adjusted.
Unless otherwise instructed, turn the pulse
rate control to the 70-120 Hz range.
Increase Button
When pressing this button, the pulse rate will
increase gradually.
Decrease Button
When pressing this button, the pulse rate will
decrease gradually.
6) Pulse Width Control
Pulse Width is adjustable. Its value is visible
on the LCD. These buttons adjusts the length
of time each electrical signal is applied through
the skin, which controls the strength and
sensation of the stimulation. If no instructions
regarding the pulse width are given in therapy,
set the control to the suggested 70-120 uS
setting.
Increase Button
When pressing this button, the pulse width will
increase gradually.
22
21
By turning the controls clockwise, the power
will be switched On. A symbol will be revealed
on the LCD as soon as the stimulation is
commenced.
The current strength of the impulses transmitted
to the electrodes increases further when the
control is turned clockwise. To reduce the
current strength or switch the device Off, turn
the control counter clockwise to the required
setting or Off-position, respectively.
3) Lead Connector
Connection of the electrodes are made with a
two-lead connector. The device must be turned
Off before connecting the cables. Both intensity
controls must be at the OFF position.
Electrodes must be pressed firmly on the skin.
4) Mode Control
Expose the controls by sliding front cover down
from middle of unit. This button operates 5
modes:
B
for burst stimulation.
M
for modulation stimulation.
C
for constant stimulation (Normal).
A
for intensity and pulse rate modulation.
S
for intensity and pulse width modulation.
Press the button until the unit is in the mode
desired.
I