TROUBLE SHOOTING
If your
GF-TX5
appears to be functioning
improperly, use the chart below to determine
what may be wrong.
LCD Indicator lights ON but unit does not
function properly:
1) Check control settings. Are they set as
prescribed by your medical professional?
2) Are electrodes in proper position?
3) Check lead wires. Make sure all connectors
are firmly seated.
4) Replace lead wires with another to check
for broken wires.
LCD light dim & Battery power indicator
shows half-scale:
Replace battery with a new one.
No LCD indicator lights on:
Replace battery with a new one.
If none of these measures correct the problem,
the unit should be serviced.
30
29
SAFETY-TECHNICAL CONTROLS
For safety reasons, check your unit each week
based on the following checklist.
1) Check the device for external damage.
Deformation of the housing.
Damaged or defective output sockets.
2) Check the device for defective operating
elements.
Legibility of inscriptions and labels.
Make sure the inscriptions and labels are
not distorted or removed.
3) Check the usability of accessories.
Patient cables are undamaged.
Electrodes are undamaged.
Please consult your dealer/ distributor if there
are any problems with device and accessories.