7.
Disconnect the white nylon tube going from the concentrator to the bottom of the regulator by carefully pushing up the
orange plastic collar and pulling the tube down.
8.
Remove the (4) 6mm hex screws surrounding the rectangular opening underneath the UAM.
9.
Push the top of the concentrator module tilting it towards the front of the UAM and with a very large flat screwdriver or
another hard flat tool separate the bottom of the concentrator from the bottom surface of the UAM’s interior.
10.
Once the concentrator is separated from the base, lift it and pull it towards you and letting it rest on the back edge of
the bottom surface of the UAM being careful not to pull on the concentrator monitor and LED indicator panel cables.
11.
Disconnect the concentrator monitor cable and LED indicator panel cable.
12.
Lift and lower the concentrator module to the floor right behind the UAM.