25
REMOVING THE CONCENTRATOR
Materials and tools required:
g
6 mm hex key
g
2.5 mm hex key
g
M5 flare nut wrench
g
Long flat screw driver or other long and flat metal tool
1.
Engage the brakes on both of the UAM’s front castors.
2.
Remove the six edge screws from the lower rear panel.
3.
Tilt the panel back and remove both hoses from the air filter by carefully pushing the orange plastic collars towards the
filter and pulling the hoses away.
4.
Rest the panel against the left side of the UAM being careful not to stress the cables connected to the panel.
5.
With the nut wrench unscrew the nut on the ground post terminal on the bottom right of the panel to remove the
ground wire connecting the panel to the concentrator.
6.
Disconnect the concentrator power switch cable (4 wires: 2 blue, 1 yellow, 1 red).
Step 5
Step 6