REPLACING THE CONCENTRATOR LOSS OF POWER ALARM BATTERY
Materials and tools required:
g
3 mm hex key
g
9 volt battery
1. Turn off the UAM and disconnect it from the mains power supply.
2. If working on a Configuration
A
UAM, loosen the hex screws on the back lower panel. On a Configuration
B
UAM, loosen
the hex screw and swing open the top rear access panel to reach the alarm battery.
3. If removing the lower panel, lift and place it so that it rests on the left side of the UAM making sure not to pull on the
cables. If opening the access panel, rotate aside.
4. At the back left side of the concentrator, or to the right inside the access panel, disconnect the battery
5. Install a new 9 volt battery. Reuse the Velcro strip from the old battery.
6. Replace the lower panel or reattach the access panel with fasteners.
BATTERY CONNECTOR
VELCRO
9V BATTERY
24