REPLACING THE ANALYZER SCREEN BATTERY
Materials and tools required:
g
2.5 mm hex key
g
3.0 mm hex key
g
Long nose pliers
1.
Remove the six screws from the edges of the lower rear panel with the 3mm hex key.
2.
Tilt the panel back and remove both tubes from the primary air filter.
3.
Rest the panel against the left side of the UAM being careful not to stress the cables connected to the panel.
4.
With the long nose pliers disconnect the positive and negative battery wires.
5.
Remove the four screws on the back of the panel that attach the battery holder.
6.
Remove battery and replace with a new one
NOTE:
Replacement batteries can be ordered from Gradian but they may be less expensive and quicker to obtain from a
local battery supplier. Battery details: YUASA NP1.2-12 12V 1.2 AH sealed lead acid battery.
7.
Assemble by following the above steps in reverse.
BATTERY HOLDER SCREWS
NEGATIVE BATTERY WIRE
ANALYZER SCREEN BATTERY
POSITIVE BATTERY WIRE
BATTERY HOLDER
28
Step 4
Step 5