High-voltage module can be switched on.
High-voltage present at the test probes during measurement
Do not touch the test probes!
The specified limit value I
MAX
has been exceeded. The current limiting
function has switched the instrument to the “standby” mode.
Test completed successfully: residual voltage less than 25 V
Continuous operation with high power
during high-voltage testing
Allow the instrument to cool down for
approximately 10 minutes
Internal temperature
monitoring is defective.
Hardware error
Send the high-voltage module to the GMC-I Service GmbH
Repair and Service Department
One of the signal lamps at the high-voltage module is defective.
Refer to chapter 11.5.2, page 67 regarding replacement.
Signal / Error Message
Meaning / Cause
Remedy