34
GMC-I Messtechnik GmbH
TEST
Select the desired test.
MENU
Open the sub-menu.
4.4.1
Configuring Test Parameters
The
discharge period
(the time during which the voltage value must
be reduced to a safe level of less than 60 V) can be set within a
range of 0 to 9 seconds. “Residual voltage less than 60 V!” is dis-
played when this threshold has been reached.
Set discharge period duration.
4.4.2
Start the Test
START
Note
The voltage test can only be started with the START key
on the test instrument.
STORE
Press briefly:
Store an event.
Press and hold:
Store an event and activate the entry field.
Up to 15 characters can be entered to the entry
field, or to the info field at the bottom left as a
description for the measurement.
See chapter 5.1.1 regarding data entry.
or
STORE
Press briefly:
Save results.
The green/red signal lamp is extinguished briefly as
acknowledgement.