64
GMC-I Messtechnik GmbH
The current value for the contact current measurement is not within
the allowable measuring range. The instrument is switched to voltage
measurement: it must be assumed that dangerous contact voltage is
present.
The measurement has been saved to memory.
Data management message:
Data transmission has been successfully completed.
PROFITEST 204HP/HV High-Voltage Module
The displayed high-voltage module has been detected.
Switch the device on with the key switch.
The displayed high-voltage module has been detected.
Switch the device on with the key switch.
Green signal lamps are not illuminated
The displayed high-voltage module has not been detected:
– Power cable not plugged in
– Key switch set to „0“
– Blown fuse
– Defective signal lamp
Connect the high-voltage module:
– Plug in power cable, see chapter 3.4, page 17
– Set key switch to “I”
– Replace fuse
– Replace signal lamp
Red signal lamps are not illuminated
– Faulty data exchange between
PROFITEST 204+
and 204HP/HV
– Defective signal lamp
– Refer to following error message: “Defective instrument!”
– Replace signal lamp, see chapter 11.5.2, page 67
Current measured during the test was capacitive to a given extent.
The larger the dimensions of the device under test, the greater the
capacitive current.
Current measured during the test was primarily ohmic.
However, it must nevertheless be assumed that leakage current is pres-
ent which, under momentary conditions, does not exceed the
selected limit value. However, this situation may change over a
period of time.
Defective instrument!
Faulty data exchange between
PROFITEST 204+
and 204HP/HV.
– Instruments mounted to one another incorrectly
– After a lamp has been replaced:
The reflectors were not re-installed
– Check connections at both instruments
– Check to see if the signal tubes are covered
– Install the reflectors
– Clean dust or contamination from the transmitter and
receiver diodes at the bottom of the basic instrument,
see chapter 11.5.1, page 67
Signal / Error Message
Meaning / Cause
Remedy