36
GMC-I Messtechnik GmbH
➭
Select the desired values for test duration, test voltage,
breaking current I
MAX
and rise time (chapter 4.5.2, page 37).
Note
Do not set I
MAX
to pulse control!
➭
Press the START key.
• The red signal lamps must light up.
• The following values are displayed at the LCD:
U
T
= – – – V
I
T
= – – – mA
The following symbol appears:
Danger: High-Voltage!
Touch
neither
the test probes
nor
the device under test
during the voltage test!
A life endangering high-voltage
of up to
5 kV
is present at
the high-voltage module test probes!
➭
Pull the trigger at the marked high-voltage pistol with switch to
its limit stop and hold depressed.
• The following values are displayed at the LCD:
U
T
= selected value
I
T
= 00.0 mA or 0.00 mA
The following symbol appears:
➭
Release the trigger (switch).
• Voltage is switched off automatically after the selected test
duration has elapsed.
c) Shutdown Functions
➭
Short circuit the two high-voltage pistols.
➭
Pull the trigger at the marked high-voltage pistol with switch to
its limit stop and hold depressed.
• The instrument is shut down immediately.
• The green signal lamps must light up, and
the red signal lamps may not light up.
• The following values are displayed at the LCD:
U
T
= XXX V
I
T
= > XXX mA
The last measured voltage value before breaking current was
exceeded is displayed along with breaking current I
MAX
.
The following symbol appears: