18
GMC-I Messtechnik GmbH
5
Controls, Display Elements and Terminals
5.1
Front Panel SYSKON P500 / P800 / P1500
*
valid as from revision level 02 and firmware version 004. In the case of hardware revision level < 02, the LED lights up yellow in both cases.
Protective Functions and Status Displays
OTP/OVP
LED lights up Overtemperature protection triggered or
output voltage
80 V exceeded,
OUTPUT = off
OVP
LED lights up
Overvoltage detection triggered,
ovset
and
ovdly
parameter values exceeded
(prerequisite:
ovp
parameter =
on
)
OUTPUT = off
OCP
LED lights up
Overcurrent detection triggered,
oCset
and
oCdly
parameters exceeded
(prerequisite:
oCp
parameter = on)
OUTPUT = off
OVP ON
LED lights up Overvoltage shutdown is activated
(
ovp
parameter in device setup menu =
on
)
OCP ON
LED lights up Overcurrent shutdown is activated
(
oCp
parameter in device setup menu =
on
)
Control Mode Status Display
CV
LED lights up
Constant voltage regulating mode: Uout
Uset
LED
Plim / CP*
lights up
yellow:
Pout > Pnom (OL)
green*:
PSET < Pnom progr. (CP)
CC
LED lights up
Constant current regulating mode: Iout
Iset
Mains Switch
POWER
LED lights up
Device is switched on
POWER
switch
Switches the device on and off
Power Output On/Off Key
OUTPUT
LED lights up Output is active
OUTPUT
switch
Switches the power output on and off
Rotary Encoder for Voltage
Voltage setpoint Uset
– adjust output voltage Uset
Condition:
UL_L (lower setting limit)
Uset
UL_H (upper setting limit)
When the rotary encoder is activated the display is switched to Uset (LED) and the cursor becomes active – the selected
decimal place blinks at the display and can be selected with the
and
scroll keys.
The new setpoint becomes effective immediately.
1
4
5
6
12
13
Rack mounting tabs
Front handles
2
Device feet
3
Eve
nts
Set
tin
gs
Display A
Standard display: measured
voltage value Uout, display
function is changed by pressing
the SELECT A key or by
turning the rotary knob. Adjust
resolution (select decimal place)
with the
and
scroll keys
Select A
Display selection:
Uout
Uset
OVset
Pset
Uset adjusting alternatives
Activate by turning the rotary encoder, or select with Select A Uset
Scroll keys
: select decimal place
: immediately increase or reduce Uset
Numeric keypad
Entry of numeric values, Uset LED blinks Execute with
, or abort with ESC
/CP