64
GMC-I Messtechnik GmbH
11
System Messages
Procedures and entries are monitored in order to provide the user
with support.
The device is capable of detecting and reading out a great variety
of defective procedures.
In the case of manual operation, the error appears directly at the
display for a brief period of time. In addition to this, the last error
message can be viewed with the help of the appropriate menu
function.
Err
appears at the left-hand display, and the three digit error code
at the right-hand display.
In the case of operation with a PC, the last three error messages
can be accessed by executing the ERROR? command (see
ERROR?
command in the section entitled “Descriptions of Oper-
ating Commands”.
Code
Err...
Meaning / Cause
Remedy
00
No error
01
TYPE (BZ) detection
For internal production process only
05
UNKNOWN KEYCODE, or in case of “LCL LOCKED”, [ESC] has been briefly
activated for enabling
12
CMD buffer overflow
Overflow at internal CMD buffer
21
USET, ISET, PSET (parameter error)
22
UL,IL (parameter error)
29
DDTE *DDT command string > 80 characters or ?*TRG? within *DDT ...
31
CME command error
General
32
EXE execution error
General
Interface
51
RS 232, PB, parity bit
52
RS 232, SB, stop bit
53
PB + SB, RS 232
54
RS232, FRAME OVERFLOW (impermissible combination DB/PB/SB)
Only occurs during manual configuration
55
Active talker state but no listener present
Only possible with IEEE 488 interface option
56
IEC$LATA$ERR (active listener and active talker)
Only possible with IEEE 488 interface option
61
ADJUST parameter error
62
ADJUST order impermissible (REMOTE)
63
[U/I]-OFFSET/FULL SCALE -> (!) [CV/CC] MODE
Corresponding operating mode required!
64
ADJUST LIMITs or OFFSET (MEASUREMENT NEGATIVE or OVERFLOW)
66
CALIBRATION ERROR/EXIT (-> UNCAL)
69
MEMORY DATA (-> ERROR) ?)
Faulty data, possible cause: low battery
71
Table values (#): USET<UL_L, USET>UL_H, ISET<IL_L, ISET>IL_H
Limit error during sequence execution
73
OUTE: “OUTPUT ON” =/= Tx MODE “OUT” & Tx-SIG “OUT OFF”
Trigger mode power output and signal blocked (=OFF!)
74
TRGE: “MINMAX ON” =/= Tx MODE “MIN” & Tx-SIG “MINMAX OFF”
Min-max control for the trigger mode and signal blocked (=OFF!)
75
TRGE: “SEQUENCE ON” =/= Tx MODE “SEQ” & Tx-SIG “SEQUENCE STOP”
Sequence control for the trigger mode and signal blocked (=OFF!)
76
TRGE: “ANALOG INP” =/= Tx MODE “AI?” & Tx-SIG (UEXT,IEXT)
Uext, Iext control for the trigger mode and signal blocked (=ON!)
81
RCL n (no data): SETUP memory n: invalid or no data
82
START -> STOP - INVALID VALUES
83
START-ADR > STOP-ADR
84
Not (STOP_ADR < MEM < START_ADR) or SUBSEQUENCE (cond.)
Address range or storage address is out of actual start-stop-address range or is an active sub-
sequence. Specify the sequence range for delete and repeat command.
85
CONTINUE (no initialization, status =/= “HALT”)
86
SUBSUBSEQUENCE impermissible
89
“SEQUENCE OFF” required for current command
91
SELFTEST (*TST?)
96
MIN LIMIT UNDERFLOW, direct entry of a numerical value
requires > UL_L and/or > IL_L
93
Current command not permissible for power control!
Impermissible combination of functions
96
MIN LIMIT UNDERFLOW
Numerical direct entry of a numerical value > UL_L and/or > IL_L necessary
97
MIN LIMIT UNDERFLOW
98
MAX LIMIT OVERFLOW
99
OVERLOAD / OVERFLOW