Terminating device messages:
The following end-of-text characters can be used for data receipt:
With IEC bus control: NL (hex: 0A) or NL & EOI or DAB & EOI;
With RS 232C control: NL or CR (hex: 0D) or ETB (hex: 17) or ETX (hex: 03).
The following end-of-message character is used when transmitting the response string:With IEC bus control: NL & EOI;
With RS 232C control: last received end-of-message character.
Abbreviating commands
: Abbreviated commands are identified with boldface letters. Letters not printed in boldface can be omitted. Example: “
OU
TPUT ?” = “OU?”
As a rule, letters can be entered in upper or lower case.
Stringing commands together
: Several commands is a single data string must be separated by semicolons (;). Example: “USET?; ISET?; OUTPUT?”
1)
Constant voltage and/or constant current mode
2)
Constant power mode
* /15 or /1700, respectively as from firmware version 004
xx: manual: division into partial functions
MI
NMAX
?
Min-Max storage for measured U and I
values
txt
OFF / ON
“ON”
enable tolerance band func-
tion for CRB.0/1, ERC.0/1, SIGx_OUT
X
X
MINMAX OFF
10
MO
DE
?
Momentary control mode of the power output txt
CV / CC / CP / OL / OFF
LED X
MODE
CV
8
OC_D
ELAY
?
Overcurrent protection triggering delay w
0.000
w
65.535 [s]
X
X
OC_DELAY 00.000
15
OCP?
Overcurrent protection
txt
OFF / ON / R01
R12/15*
X
X
OCP OFF
7
OCS
ET
?
Overcurrent protection trigger value
w
+XXX.XXX [A]
X
X
OCSET +080.000
15
OU
TPUT
?
Output on-off status
txt
OFF/ ON
LED X
OUTPUT
ON
10
OV_D
ELAY
?
Overvoltage protection triggering delay
w
XX.XXX [s] 0
w
65,535 [s]
X
X
OV_DELAY 00.000
15
OVP?
Overvoltage protection
txt
OFF / ON / R01
R12/15*
X
X
OVP
ON
7
OVS
ET
?
Overvoltage protection trigger value
w
+XXX.XXX [V]
X
X
OVSET +080.000
15
POU
T
?
Current output power
w
+XXXXX.X [W]
X
X
POUT +00002.1
14
POW
ER_ON
?
Response after power on
txt
RST / SBY / RCL / R01
R12/15*
X
X
POWER_ON SBY
12
PS
ET?
Power setpoint [W]
w
+XXXXX.X [W]
W = Pnom (Pnom/2 for 115 Vac)
1)
W
Pnom (Pset<Pnom/2 for 115 Vac)
2)
X
X
PSET +01500.0
13
RE
PETITION
?
(i)
Number of repetitions for sequence
function
i is an optional parameter for the setup
memory location (1 through 12/15*),
which should be read out directly.
n
0
n
255;
0 means continuous repetition
X
X
REPETITION 000
15
RL
OAD?
Momentary load resistance
[Calculated value R = U/I]
w
+XXX.XXX [
]
X
X
RLOAD +030.833
14
SEQ
UENCE
?
SEQUENCE Status
txt,n1,n2,n3 txt: RDY / HALT / RUN
n1: 000
n1
012/15* (setup memory)
n2: 001
n2
255 (remaining
repetitions),
n2 = 999 = continuous
n3: 0001
n3
1536/1700*
(memory location)
xx X
SEQUENCE
RDY,000.999,0005
26
SIG
123
?
Analog interface signal outputs
txt1,txt2,txt3 OFF / ON / OUT / MODE / SEQ /
SSET / U_LO / U_HI / I_LO / I_HI
X
X
SIG123 MODE, OUT, OFF
21
SINK?
Sink function on/off
txt
OFF / ON
X
X
SINK
ON
8
SS
ET
?
Status of an assigned switching function txt
OFF / ON
X
X
SSET OFF
8
STA
RT_STOP
?
(i)
Start and stop address
i is an optional parameter for the setup
memory location (1 through 12/15),
which should be read out directly.
m,n
1
n1
n2
1536/1700*
xx X
START_STOP 0001.0005
20
STO
RE
?
(m(,n(,tab)))
Read data from sequence memory
See “Descriptions of Operating
Commands” for further details.
n,w1,w2,w3,
txt
n: memory location address
w1: +XXX.XXX [V]
w2: +XXX.XXX [A]
w3: XX.XXX [s]
txt: “FSET”
xx X
STORE 0003,+020.000,+015.000,0
0.000,
NF
40
T_M
ODE
?
Function status for trigger inputs 1 and
2
txt1,txt2
OFF / OUT / SQS / SEQ(,n) / LLO /
MIN (,n) / AIX / AIU / AII
X
X
T_MODE OUT,LLO
15
TD
EF
?
(i)
Default time for sequence function
i is an optional parameter for setup
memory (1 through 12/15), which
should be read out directly.
w
0.001
w
65.535[ s]
X
X
TDEF 01.000
12
TIM
EDATE
?
System clock time/date (RTC)
txt
yyyy-mm-ddThh:mm:ss
xx X
TIMEDATE 2007-10-08T12:27:13
28
TS
ET
?
Memory location-specific dwell time
for sequence function
w
0.000
w
65.535[ s]
0.000 = Tdef,
X
X
TSET 00.000
12
UI
_C_SET
?
Reference values for Uout/Iout,
tolerance band function
w1,w2,w3,w
4
[V][V][A][A]
w1, w2: 0
w1 < w2
Unom [V]
w3, w4: 0
w3 < w4
Inom [A]
xx X
UI000.000,+060.000,+00
0.000,+060.000
44
UL_H?
(
ULI
M
?
)
Upper limit value for voltage setting
w
+XXX.XXX [V]
X
X
UL_H +060.000
13
UL_L?
Lower limit value for voltage setting
w
+XXX.XXX [V]
X
X
UL_L +000.000
13
UMA
X
?
Max. measured voltage value
w
+XXX.XXX [V]
X
X
UMAX +010.004
13
UMI
N
?
Minimum measured voltage value
w
+XXX.XXX [V]
X
X
UMIN +009,992
13
UO
UT
?
Presently measured voltage value
w
+XXX.XXX [V]
X
X
UOUT +009.998
13
US
ET
?
Selected voltage setpoint
w
+XXX.XXX [V]
X
X
USET +010.000
13
Query Command
Meaning
Response
Parameter
Values / Format
Manual
Re
mote
Sample Response
Response
String